4.1 Comparator Function
53
4
Chapt
er
4
Appl
ie
d Measur
ement
To enter the current measurement as the setting value:
AUTO
key
Press on a screen other than the upper/lower threshold (reference value/toler-
ance) setting display. This key is used as a numeric key on the upper/lower
threshold (reference value/tolerance) setting display.
The current measurement value is set as the upper or lower threshold (during
upper/lower threshold setting), or as the reference value (during reference value
and tolerance setting). If the measured value is faulty or
±
OF, it is ignored (not
entered).
To enter a statistical calculation result as the setting value:
STAT
key
Press on a screen other than the upper/lower threshold (reference value/toler-
ance) setting display. This key is used as a numeric key on the upper/lower
threshold (reference value/tolerance) setting display.
The result of statistical calculation is set as follows:
Where “
” represents population standard deviation (
n
).
No setting occurs if statistical calculation is disabled and no statistical calculation
result exists.
See
"4.4 Statistical Calculation Functions" (
Setting thresholds from the
AUTO
and
STAT
keys is possible only when the
selected (blinking) character is non-numeric.
This section describes how to configure functionality for acquiring the absolute
value of the measured voltage value when judging comparators, allowing a judg-
ment to be made based on the absolute value of the voltage even if polarity is
reversed when the probes are connected to the battery.
Ordinarily, connecting the probes with the polarity reversed results in a negative
measured voltage value, yielding a Lo comparator judgment result. To generate
an IN judgment whenever the reading falls within the specified range, even if the
probes have been connected backwards (resulting in a negative voltage mea-
sured value), set the absolute value judgment function to "On."
This function is configured on the menu screen.
During upper/lower
threshold setting
Upper threshold = average value + 3
Lower threshold = average value - 3
During reference value
and tolerance setting
Reference value = average value
Tolerance = 3
/ average value
×
100%
Threshold and reference values can be set from 0 to 99999 (or 999999 for volt-
age), and tolerance can be set from 0.000 to 99.999%. Negative values are not
settable. Entries using statistical calculation results that exceed the valid range
are restricted to the range limit.
Configuring the Absolute Value Judgment Function (Voltage)
1
(The SHIFT indicator lights up.)
The menu screen is displayed.
HIOKI BT3564A961-02
Summary of Contents for BT3564
Page 2: ...HIOKI BT3564A961 02...
Page 44: ...3 7 Displaying Measurement Results 38 HIOKI BT3564A961 02...
Page 80: ...4 12 Reset Function 74 HIOKI BT3564A961 02...
Page 98: ...6 3 Printing 92 HIOKI BT3564A961 02...
Page 170: ...8 8 Sample Programs 164 HIOKI BT3564A961 02...
Page 182: ...10 3 Error Indication 176 HIOKI BT3564A961 02...
Page 202: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3564A961 02...
Page 206: ...Index 4 HIOKI BT3564A961 02...
Page 207: ...HIOKI BT3564A961 02...
Page 208: ...HIOKI BT3564A961 02...
Page 209: ...HIOKI BT3564A961 02...
Page 210: ...HIOKI BT3564A961 02...