3.5 Setting Sampling Rate
30
The sampling rate can be selected from FAST, MEDIUM, or SLOW. Slower
sampling rates generally provide greater measurement precision.
Execute zero adjustment before measuring to nullify any residual offset voltage
from the instrument or measurement environment. Measurement accuracy spec-
ifications are applicable after zero adjustment. Zero adjustment can also be exe-
cuted by the 0ADJ terminal of the EXT I/O connector.
See
Before executing zero adjustment, connect the test leads (probes) as follows:
1. Connect SENSE-H to SENSE-L.
2. Connect SOURCE-H to SOURCE-L.
3. Connect the joined SENSE and SOURCE leads together as shown below.
3.5 Setting Sampling Rate
Selects the sampling rate
FAST
MEDIUM
SLOW
• When SLOW sampling rate is selected, self-calibration is executed during
each measurement. At other sampling rates, self-calibration is executed
manually or automatically every 30 minutes.
See
• Refer to the specifications for details of sampling time.
See
3.6 Zero-Adjust Function
Wiring Method for Zero-Adjustment
Connection
Connection
SENSE-H
SENSE-L
SOURCE-H
SOURCE-L
HIOKI BT3564A961-02
Summary of Contents for BT3564
Page 2: ...HIOKI BT3564A961 02...
Page 44: ...3 7 Displaying Measurement Results 38 HIOKI BT3564A961 02...
Page 80: ...4 12 Reset Function 74 HIOKI BT3564A961 02...
Page 98: ...6 3 Printing 92 HIOKI BT3564A961 02...
Page 170: ...8 8 Sample Programs 164 HIOKI BT3564A961 02...
Page 182: ...10 3 Error Indication 176 HIOKI BT3564A961 02...
Page 202: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3564A961 02...
Page 206: ...Index 4 HIOKI BT3564A961 02...
Page 207: ...HIOKI BT3564A961 02...
Page 208: ...HIOKI BT3564A961 02...
Page 209: ...HIOKI BT3564A961 02...
Page 210: ...HIOKI BT3564A961 02...