8.6 Message Reference
144
EXT I/O Input
Syntax
Query
:IO:IN?
Response
0 to 31(NR1)
Description
Query
Signals at the EXT I/O (IN0 to IN4) input terminals are read at the
leading edge.
Each bit (edge data) is cleared upon reading by this query.
A bit is set when the leading edge (short-circuit between each signal
terminal and the ISO_COM terminal) is detected, and is cleared when
read by this query command.
See
Note
The
TRIG
key and
TRG
command are detected in the same way as the TRIG
terminal signal.
bit4
bit3
bit2
bit1
bit0
IN4 (MA-
NU)
IN3
(PRINT)
IN2 (0ADJ)
IN1 (CAL)
IN0 (TRIG)
Pin No.
26
7
21
20
1