8.6 Message Reference
121
8
Chapt
er 8
RS-232
C
/GP-I
B
In
te
rf
aces
Set and Query the Standard Event Status Enable Register (SESER)
Syntax
Command
ESE
<0 to 255>
Query
ESE?
Response
<0 to 255 (NR1)>
Description
Command The SESER mask is set to the numerical value 0 to 255.
The initial value (at power-on) is 0.
Query
The contents of the SESER, as set by the
ESE
command, are
returned as an NR1 value (0 to 255).
Example
Command
ESE 36
Sets bits 5 and 2 of SESER.
Query
ESE?
Response
36
SESER has been set to bit 5 and bit 2.
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
PON
URQ
CME
EXE
DDE
QYE
RQC
OPC
Query and Clear the Standard Event Status Register (SESR)
Syntax
Query
ESR?
Response
<0 to 255 (NR1)>
Description
Query
Returns the contents of the SESR as an NR1 value from 0 to 255,
then clears register contents.
The response message has no header.
Example
Query
ESR?
Response
32
Bit 5 of the SESR was set to 1.
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
PON
unused
CME
EXE
DDE
QYE
unused
unused
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
PON
URQ
CME
EXE
DDE
QYE
RQC
OPC
HIOKI BT3564A961-02
Summary of Contents for BT3564
Page 2: ...HIOKI BT3564A961 02...
Page 44: ...3 7 Displaying Measurement Results 38 HIOKI BT3564A961 02...
Page 80: ...4 12 Reset Function 74 HIOKI BT3564A961 02...
Page 98: ...6 3 Printing 92 HIOKI BT3564A961 02...
Page 170: ...8 8 Sample Programs 164 HIOKI BT3564A961 02...
Page 182: ...10 3 Error Indication 176 HIOKI BT3564A961 02...
Page 202: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3564A961 02...
Page 206: ...Index 4 HIOKI BT3564A961 02...
Page 207: ...HIOKI BT3564A961 02...
Page 208: ...HIOKI BT3564A961 02...
Page 209: ...HIOKI BT3564A961 02...
Page 210: ...HIOKI BT3564A961 02...