8.6 Message Reference
119
8
Chapt
er 8
RS-232
C
/GP-I
B
In
te
rf
aces
Messages specific to the RS-232C or GP-IB interface are identified by their corre-
sponding symbols.
System Data Command
Internal Operation Command
Standard Commands
Queries device ID.
Syntax
Query
IDN?
Response
<Manufacturer's name>,<Model name>,0,<Software version>
Description
Query
Queries the device manufacturer's name, model name and software
version.
Example
Query
IDN?
Response
HIOKI,BT3564,0,V1.00
The Device ID is Hioki BT3564, 0, software version 1.00.
Note
• The response message has no header.
Initialize Device
Syntax
Command
RST
Description
Command Resets instrument settings (other than saved data) to factory defaults.
Operation returns to the initial display after initialization.
Note
• The communication conditions are not initialized.
• To initialize saved data as well, send the
:SYSTem:RESet
command.
Execute Self-Test and Query the Result
Syntax
Query
TST?
Response
<0 to 3>
0....... No Errors
1....... RAM Error
2....... EEPROM Error
3....... RAM and EEPROM Errors
Description
Query
Perform instrument self-test and return the result as numerical value 0
to 3.
Example
Query
TST?
Response
1
A RAM Error occurred.
HIOKI BT3564A961-02
Summary of Contents for BT3564
Page 2: ...HIOKI BT3564A961 02...
Page 44: ...3 7 Displaying Measurement Results 38 HIOKI BT3564A961 02...
Page 80: ...4 12 Reset Function 74 HIOKI BT3564A961 02...
Page 98: ...6 3 Printing 92 HIOKI BT3564A961 02...
Page 170: ...8 8 Sample Programs 164 HIOKI BT3564A961 02...
Page 182: ...10 3 Error Indication 176 HIOKI BT3564A961 02...
Page 202: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3564A961 02...
Page 206: ...Index 4 HIOKI BT3564A961 02...
Page 207: ...HIOKI BT3564A961 02...
Page 208: ...HIOKI BT3564A961 02...
Page 209: ...HIOKI BT3564A961 02...
Page 210: ...HIOKI BT3564A961 02...