Hioki 9695-02 Instruction Manual Download Page 1

Introduction

Thank you for purchasing the HIOKI 9695-02/03 CLAMP ON SEN-
SOR. To obtain maximum performance from the product, please read 
this manual first, and keep it handy for future reference.

Overview

The 9695-02/03 are voltage-output, clamp-on sensors used for 50 A
(9695-02) and 100 A (9695-03) rated AC measurement.
The 9695-02/03 sensors are designed to measure alternating currents
on a hot conductor without disconnecting it from the power line. Both
sensors are easy to operate and suitable for electric current and
power measurement in various fields.

Initial Inspection

When you receive the product, inspect it carefully to ensure that no
damage occurred during shipping. If damage is evident, or if it fails to
operate according to the specifications, contact your dealer or Hioki
representative.

Maintenance and Service

• To clean the product, wipe it gently with a soft cloth moistened with

water or mild detergent. Never use solvents such as benzene, alco-
hol, acetone, ether, ketones, thinners or gasoline, as they can
deform and discolor the case.

• If the product seems to be malfunctioning, contact your dealer or

Hioki representative. Pack the product carefully so that it will not be
damaged during shipment, and include a detailed written description
of the problem. Hioki cannot be responsible for damage that occurs
during shipment.

(Accuracy guaranteed for one year at 23°C±5°C(73°F±9°F), 80%RH or less. 
Endurance number of the 

jaw

 opening and closing part: 10000 times)

f.s. : maximum display value or scale length
rdg.: reading value (The value currently being measured and indicated on the 

measuring product)

Follow these precautions to ensure safe operation and to obtain the
full benefits of the various functions.

Measurement categories 

This product conforms to the safety requirements for CAT III measurement 
products.
To ensure safe operation of measurement products, IEC 61010 establishes 
safety standards for various electrical environments, categorized as CAT II to 
CAT IV, and called measurement categories. 
CAT II: Primary electrical circuits
in equipment connected to an AC
electrical outlet by a power cord
(portable tools, household appli-
ances, etc.)
CAT II covers directly measuring
electrical outlet receptacles.
CAT III: Primary electrical cir-
cuits of heavy equipment (fixed
installations) connected directly
to the distribution panel, and
feeders from the distribution panel to outlets.
CAT IV: The circuit from the service drop to the service entrance, and to the
power meter and primary overcurrent protection device (distribution panel).

Using a measurement product in an environment designated with a higher-
numbered category than that for which the product is rated could result in a 
severe accident, and must be carefully avoided.
Use of a measurement instrument that is not CAT-rated in CAT II to CAT IV 
measurement applications could result in a severe accident, and must be care-
fully avoided.

Safety Symbol

The following symbols in this manual indicate the relative importance

of cautions and warnings.

This manual contains information and warnings essential for safe
operation of the product and for maintaining it in safe operating condi-
tion.  Before using the product, be sure to carefully read the following
safety notes.

• Accurate measurement may be impossible in the presence of strong

magnetic fields, such as near transformers and high-current conduc-
tors, or in the presence of strong electromagnetic fields such as near
radio transmitters.

• When connecting the cables to the output terminal, twist the cables

together as close to the terminal as possible to avoid the influence of
an external magnetic field 

• Use the 9219 CONNECTION CABLE to connect this product to the

HIOKI clamp wattmeter. (The connector of the 9219 is a solderless
terminal to BNC.)

Preliminary Checks

Before using the product the first time, verify that it operates normally
to ensure that the no damage occurred during storage or shipping. If
you find any damage, contact your dealer or Hioki representative.

Attach the clamp around only one conductor. Single-phase (2-wire) or
three-phase (3-wire) cables clamped together will not produce any reading.

9695-02, 9695-03

CLAMP ON SENSOR

Instruction Manual

July 2015  Revised edition 8  Printed in Japan

9695B980-08  15-07H

Warranty

Warranty malfunctions occurring under conditions of normal use in
conformity with the Instruction Manual and Product Precautionary
Markings will be repaired free of charge. This warranty is valid for a
period of one (1) year from the date of purchase.  Please contact
the distributor from which you purchased the product for further
information on warranty provisions.

Specifications 

Rated primary current

9695-02: 50 A AC, 9695-03: 100 A AC

Output voltage

9695-02: 10 mV AC/A, 9695-03: 1 mV AC/A

Amplitude accuracy

±

0.3% rdg.±0.02%f.s. (f.s.:50 A (9695-02)/ 100 A (9695-03), 

45Hz to 66Hz, at 

jaw

 center)

Phase accuracy

9695-02: Within 

±

2° (45 Hz to 5 kHz)

9695-03: Within 

±

1° (45 Hz to 5 kHz)

Amplitude frequency 
characteristics

Within ±1% at 40 Hz to 5 kHz (deviation from accuracy)

Effect of conductor 
position

Within ±0.5% (deviation from center)

Effect of external 
electromagnetic field

0.1 A equivalent or lower
(in an AC electromagnetic field of 400 A/m)

Maximum permissible 
input

60 A continuous (9695-02)/ 130 A continuous (9695-03)
(at 45 to 66 Hz, ambient temperature 50

°

C)

Temperature coefficient

0.02%rdg./

°

C

Dielectric strength

3536 V ACrms for 15 sec (between electric circuit and 

jaw

)

Maximum rated voltage to 
earth

300 V ACrms or lower (Insulated conductor)

Operating Temperature 
&Humidity

0 to 50°C (32 to 122°F), 80%RH or lower 
(non-condensating)

Storage Temperature 
&Humidity

-10 to 60°C(14 to 140°F), 80%RH or lower 
(non-condensating)

Operating Environment

Indoors, up to 2000 m (6562-ft.) ASL

Standards applying

Safety

EN61010
Measurement Category III, Pollution Degree 2
(Anticipated Transient Overvoltage: 4000 V)

EMC EN61326

Measurable conductor 
diameter

15 mm (0.59”) or less

Output terminals

M3 terminal (Maximum outside diameter: 6.5 mm/0.26”, inside 
diameter: 3.2 mm/0.13”)

Cable length

3 m or less

Dimensions

Approx. 50.5W x 58.0H x 18.7D mm (1.99”W x 2.28”H x 
0.74”D) (excluding protrusions)

Mass

Approx. 50 g (1.8 oz.)

Accessory

Instruction Manual

Option

9219 CONNECTION CABLE, 9238 CLAMP SENSOR CABLE

Safety

This product is designed to conform to IEC 61010 Safety Stan-
dards, and has been thoroughly tested for safety prior to ship-
ment. However, mishandling during use could result in injury or
death, as well as damage to the product. Be certain that you
understand the instructions and precautions in the manual
before use. We disclaim any responsibility for accidents or inju-
ries not resulting directly from product defects.

Indicates cautions and hazards. When the symbol is printed on the 

instrument refer to a corresponding topic in the Instruction Manual.

Indicates a double-insulated device.

Indicates AC (Alternating Current).

Wear appropriate protective insulation (insulating rubber gloves 

and boots, helmet and etc.) when connecting and disconnecting 

from live electric circuits.

Indicates that incorrect operation presents an extreme hazard that 

could result in serious injury or death to the user.

Indicates that incorrect operation presents a significant hazard 

that could result in serious injury or death to the user.

I

ndicates that incorrect operation presents a possibility of injury to 

the user or damage to the product.

Advisory items related to performance or correct operation of the 
product.

Usage Notes 

• To avoid short circuits and potentially life-threatening

hazards, never attach the product to a circuit that oper-
ates at more than the 300 V, or over bare conductors.

• This product should only be connected to the sec-

ondary side of a breaker, so the breaker can prevent
an accident if a short circuit occurs. Connections
should never be made to the primary side of a
breaker, because unrestricted current flow could
cause a serious accident if a short circuit occurs.

• To avoid electric shock, do not allow the product to get wet,

and do not use it when your hands are wet.

• To avoid electric shock when measuring live lines, wear

appropriate protective gear, such as insulated rubber gloves,
boots and a safety helmet.

• Note that the product may be damaged if the applied current ex-

ceeds the measurement range.

• Do not store or use the product where it could be exposed to direct

sunlight, high temperature or humidity, or condensation. Under
such conditions, the product may be damaged and insulation may
deteriorate so that it no longer meets specifications.

• Be careful to avoid dropping the clamps or otherwise subjecting

them to mechanical shock, which could damage the mating surfac-
es of the core and adversely affect measurement.

• Keep the jaw free from foreign objects, which could interfere with

clamping action.

• Keep the jaw closed when not in use, to avoid accumulating dust

or dirt on the mating jaw, which could interfere with clamp perfor-
mance.

• Do not fix the output terminal screws too tightly. The torque about

0.5N

m is recommended.

Before using the product, make sure that no bare conductors
are improperly exposed.  Using the product in such conditions
could cause an electric shock, so contact your dealer or Hioki
representative for repair.

Parts Names

Measurement Procedures 

• Do not disconnect the output cable when the tester connected to

the clamp-on sensor is on or with the sensor clamped on a conduc-
tor to measure. This prevents the 9695-02/03 main unit and sensor
from malfunctioning.

• M3x5 screws and a spring washer assembly should be used for the

9695-02/03 terminals. When using another type of screw (especial-
ly longer ones), be careful with the tightening torque. Tightening a
screw too much may damage the 9695-02/03.

Output

Current direction indicator

Current direction indicator

   

Jaw

Lever

Lock knob

terminals

Cable polarity display

Output
terminal
cover

OK

1. Confirm that the tester con-

nected to the clamp-on sen-
sor is powered off.

2. Open the terminal cover and

connect the cable to the ter-
minal of the 9695-02/03.
Make sure that the polarities
match.

3. Connect the cable to the

input terminal of the tester.

4. Turn on the tester.

5. If the lever is locked, unlock

it.

6. Open the jaw. Orient the cur-

rent direction indicator to the
load side and clamp one con-
ductor at the center of the
jaw.

7. Close the jaw and lock it. Be

sure to lock the clamp core
since it has no spring.

LO

AD

SO

UR

CE

Position the clamp with the current 

direction indicator pointing toward the 

load side. (If installed in the opposite 

direction,the phase deviates 180 

degrees.)

Current direction 

indicator

Clamp the conductor.

Electric 

conductor

Unlock the lever

1

2

When output cover 

is opened