HEADQUARTERS:
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 FAX +81-268-28-0568
http://www.hioki.com / E-mail: [email protected]
HIOKI USA CORPORATION:
TEL +1-609-409-9109 FAX +1-609-409-9108
http://www.hiokiusa.com / E-mail: [email protected]
All information correct as of Dec. 27, 2012. All specifications are subject to change without notice.
DMMsE
8
-
36
E Printed in Japan
HIOKI (Shanghai) SALES & TRADING CO., LTD.:
TEL +86-21-63910090 FAX +86-21-63910360
http://www.hioki.cn / E-mail: [email protected]
HIOKI INDIA PRIVATE LIMITED:
TEL +91-731-6548081 FAX +91-731-4020083
E-mail: [email protected]
HIOKI SINGAPORE PTE. LTD.:
TEL +65-6634-7677 FAX +65-6634-7477
E-mail: [email protected]
HIOKI KOREA CO., LTD.:
TEL +82-42-936-1281 FAX +82-42-936-1284
E-mail: [email protected]
РАДАР - ОФИЦИАЛЬНЫЙ ДИЛЕР HIOKI
РОССИЯ, 198152, Санкт-Петербург
Краснопутиловская ул., д.25
Тел./факс +7 (812) 600-48-89
Тел.: +7 (812) 375-32-44
www.radar1.ru
OVERVIEW
Easy-to-use, current to current transformer
AC current up to 1500 A, secondary current 1/10 of primary
Basic specifications
(Accuracy guaranteed for 1 year)
AC 1000 A continued (Maximum 1500 A for 5 minutes or shorter)
Rated primary current
Rated secondary current
Amplitude Accuracy
Phase Accuracy
AC 100 A (10 : 1 CT ratio)
±1.5% rdg.
±1.0° or less
Frequency characteristic
Amplitude: 20 Hz to 5 kHz: ±2.0 % rdg. (deviation from accuracy)
Phase: 20 Hz to 5 kHz: ±1.0° or less (deviation from accuracy)
Max. rated voltage to earth
Core jaw dia.
Dimensions, mass
Supplied accessories
600 V AC rms (insulated wire)
φ55 mm (2.17 in) or 80 mm (3.15 in) × 20 mm (0.79 in) busbar
99.5 mm (3.92 in) W × 188 mm (7.40 in) H × 42 mm (1.65 in) D, 580 g (20.5 oz), cord
length 3 m (9.84 ft)
Instruction manual ×1, Mark band ×6
Note: Cannot use with the Model CT 9279
Summary of Contents for 9290-10