14.2 External I/O
380
You can output a signal when a trigger event occurs.
You can also connect several instruments for parallel trigger synchronized oper-
ation.
See
"Parallel Trigger Synchronization" (p. 379)
14.2.2 Trigger Output
(TRIG OUT/CAL)
Pulse wave is output.
Trigger event occurs
TRIG OUT/CAL
The TRIG OUT/CAL terminal can be used as a trigger output terminal (TRIG
OUT) or as a probe calibration output signal terminal (CAL) (p. 393). It cannot be
used for both functions at the same time.
Trigger Output Signals
Output signal
Open collector output (with voltage output), active LOW
Output voltage range
HIGH level: 4.0 to 5.0 V, LOW level: 0 to 0.5 V
Pulse width
LOW level: 1 ms or greater
Maximum input voltage
-20 to +30 V 50 mA max 200 mW max
HIGH
4 to 5.0 V
LOW
0 to 0.5 V
GND
4.7 k
TRIG OUT
4.7 k
1 k
5 V
1ms or greater
Trigger events occur and signals are output when the auto-ranging function
(
FUNCTION MODE
F4
[Auto Setting]
) (p. 74) is used with the Memory Function.
You should be aware of this if you are using the trigger output terminal together
with the auto-ranging function.
Summary of Contents for 8860-50
Page 2: ......
Page 12: ...Contents x...
Page 18: ...Reading this Manual 6...
Page 24: ...1 3 Interconnection and Block Diagrams 12...
Page 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Page 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Page 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Page 406: ...14 2 External I O 394...
Page 489: ......
Page 490: ......