14.2 External I/O
378
You can input external signals as trigger sources.
When you are using several instruments, you can also synchronize triggers
(p. 379).
1
Connect the cables for the corresponding external input signals to the EXT TRIG
and GND terminals.
2
In the Trigger Settings screen, set External trigger to
[On]
14.2 External I/O
14.2.1 External Trigger Input
(EXT TRIG)
Trigger Input Signals
Voltage range
HIGH level: 3.0 to 5.0 V, LOW level: 0 to 0.8 V
Pulse width
HIGH level: 50 ns or greater, LOW level: 50 ns or greater
Maximum input voltage
-2 to 7 V
Signal Input Procedure
External signal input
Trigger event occurs.
EXT TRIG
HIGH
3.0 to 5.0 V
LOW
0 to 0.8 V
50
ns or greater
50 ns or greater
GND
150 pF
100
3.3 k
5 V
EXT TRIG
[
]
[
]
[
]: Rising setting
[
]: Falling setting
[
]
[
]
5 V
GND is common. It can be connected to any ground.
Connection procedure: