7.1 Making Input Waveform Display Settings (Analog Waveforms)
172
Set the waveform zero position (in this example, zero volts) for display on the
vertical axis.
The waveform display range can be verified on the Level Monitor.
The following two setting methods are available:
• Using the operating keys
• Using the
RANGE/POSN
knobs (settable regardless of cursor position)
7.1.2 Setting the Waveform Display Position
(Zero Position)
Normal Waveform
Changed Zero Position
100%
0V
0%
50%
Waveform
Display Range
Zero Position: 50%
0V
Zero Position: 25%
25%
Setting the Zero Position
To open the screen: Press the
SET
key
Select
with the
SUB MENU
keys
Channel Settings screen
See
To set from the Waveform screen (p. 134), To set in the Channel List (p. 130)
Using the Operating Keys
Operating Key
Procedure
1
SHEET/PAGE
Select the
[One Ch]
page.
2
CURSOR
Move the cursor to the
[Position]
item.
3
F1 to F8
Set the zero position.
The valid setting range depends on display mag-
nification.
With x 1 magnification: –100 to 150%
See
Using the RANGE/POSN Knobs
Turn the outer
POSN
knob.
Higher Position
Lower Position
Summary of Contents for 8860-50
Page 2: ......
Page 12: ...Contents x...
Page 18: ...Reading this Manual 6...
Page 24: ...1 3 Interconnection and Block Diagrams 12...
Page 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Page 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Page 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Page 406: ...14 2 External I O 394...
Page 489: ......
Page 490: ......