8.9 Message Reference
183
8
Setting and Query of Upper Limit and Lower Limit Values of First Parame-
ter for BIN Function in Deviation Percent Mode
(Only for Models 3504-50, 3504-60)
Syntax
Command
Query
Response
:BIN:FLIMit:DEViation
<BIN Number>,<Lower Limit
Value>,<Upper Limit Value>
:BIN:FLIMit:DEViation?
<BIN Number>
<BIN Number>,<Lower Limit Value>,<Upper Limit Value>
<BIN Number>
= 1 to 14 (NR1)
<Lower Limit Value>
= OFF/ Numeric Value from -999.99 to 999.99 (NR2)
<Upper Limit Value>
= OFF/ Numeric Value from -999.99 to 999.99 (NR2)
Explanation
Command Sets the upper limit and lower limit values of the first parameter in
deviation percent mode of the specified BIN number.
A numeric value in NRf format is accepted but non significant digits
are rounded off so the numeric value can be handled.
Query
Returns the upper limit and lower limit value settings for the first
parameter in deviation percent mode of the specified BIN number in
order of BIN number, lower limit value, and upper limit value.
Example
Command
:BIN:FLIMit:DEViation 1,-10.0,10.0
Sets -10 for the lower limit value and 10 for the upper limit value of the
first parameter in deviation percent mode of BIN1.
Query
Response
:BIN:FLIMit:DEViation? 1
:BIN:FLIMit:DEVIATION 1,-10.000,10.000
(when
HEADER ON)
1,-10.000,10.000
(when HEADER OFF)
-10% is set for the lower limit value and 10% is set for the upper limit
value of the first parameter in deviation percent mode of BIN1.
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......