8.9 Message Reference
175
8
(4) Status and Event Control Commands
Clearing of Status Byte Register and Related Queues (Except Output Queue)
Syntax
Command
*
CLS
Explanation
Clears the content of the event registers (SESR, ESR0, ESR1, ESR2, ESR3).
Note
The output queue is not affected
The output queue and the MAV (bit 4) of each type of enable register
status byte are not affected.
Reading and Writing of Standard Event Status Enable Register (SESER)
Syntax
Command
Query
Response
*
ESE
<Numeric Value>
*
ESE?
<Numeric Value>
<Numeric Value>
= 0 to 255 (NR1)
Explanation
Command Sets the mask pattern of the SESER to a numeric value from 0 to 255.
A numeric value in NRf format is accepted but non significant digits
are rounded off so the numeric.
The initial value (when the power is turned on) is 0.
Query
Returns the SESER content set by the ESE command as an NR1
numeric value from 0 to 255.
Example
Command
Query
Response
*
ESE 36
Sets bit 5 and bit 2 of SESER
*
ESE?
*
ESE 36
(when HEADER ON)
36
(when HEADER OFF)
Bit 5 and bit 2 of SESER are 1.
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
PON
URQ
CME
EXE
DDE
QYE
RQC
OPC
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......