7.1 About the EXT I/O Connector
123
9
7
The EXT I/O connector includes the following functions.
• Output signal for comparator result
• Output signal for BIN result (Model 3504-50, 3504-60 only)
• Output end of measurement signal (EOM)
• Output analog end of measurement signal (INDEX)
• Input external trigger signal
• Select the panel number to load
Connector Used
DDK 57RE-40500-730B (D29)
Applicable Connector DDK 57-30500
EXT I/O Connector Terminal
*The output of BIN results is only for Model 3504-50, 3504-60.
EXT I/O
Chapter 7
7.1 About the EXT I/O Connector
25
1
50
26
PIN No.
I/O
Signal Line
Name
PIN No.
I/O
Signal Line
Name
1
IN
TRIG
26
IN
LD0
2
IN
LD1
27
IN
LD2
3
IN
LD3
28
IN
LD4
4
IN
LD5
29
IN
LD6
5
IN
LD-VALID
30
OUT
BIN1, C-HI
6
OUT
BIN2, C-IN
31
OUT
BIN3, C-LO
7
OUT
BIN4, D-HI
32
OUT
BIN5, D-IN
8
OUT
BIN6, D-LO
33
OUT
BIN7, AND
9
OUT
BIN8
34
OUT
BIN9
10
OUT
BIN10
35
OUT
BIN11
11
OUT
BIN12
36
OUT
BIN13
12
OUT
BIN14
37
OUT
OUT OF BINS
13
OUT
INDEX
38
OUT
EOM
14
OUT
CV-ERR
39
OUT
D-NG
15
IN
CALIB
40
-
Unused
16 to 20
IN
EXT DCV
41 to 45
OUT
INT DCV
21 to 25
IN
EXT COM
46 to 50
OUT
INT COM
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......