6.11 Performing a System Reset
117
9
6
System reset resets the system to the factory default state, except for the set-
tings below.
• Open, short, load, offset compensation
• Keylock function
• Panel save
1.
Turn the power of the unit off.
2.
Turn the power back on while holding down
, and then let go
of
when the version information is displayed.
The version information is displayed for approximately 1.5 seconds after all
LEDs light up. Then, the system reset setting is displayed.
3.
Use the arrow keys to set system reset while the indication is dis-
played.
4.
Press
to confirm the setting.
If you execute a system reset, the measurement conditions are reset and the
system is placed in the normal measurement mode.
See
Appendix 7 "Initial Settings Table" (p. 11)
6.11 Performing a System Reset
System reset setting item
(MAIN display area)
“no”
A system reset is not performed.
“yES”
A system reset is performed.
The settings of the RS-232C interface in the 3504-40, 3504-50, 3504-60 unit
are initialized to 9600 bps for the baud rate and CR+LF for the terminator.
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......