5.4 Starting and Stopping Recording
88
5.4
Starting and Stopping Recording
To start and stop
recording
Recording can be started and stopped either manually or at
preset times.
• Manual recording
Press the
START/STOP
key to start and stop recording.
• Timed recording
After setting start and stop times, recording starts and
stops at the preset times.
With either method, always press the
START/STOP
key to
start recording. You can start and stop recording by press-
ing
START/STOP
with any screen displayed.
When recording starts, data is recorded to internal memory.
When recording with
partitioned memory
To erase recorded data
Press the
DATA RESET
key.
• After pressing the
DATA RESET
key, execute "Store
Recording Data".
• After pressing the
DATA RESET
key, execute "Erase the
Data Just Recorded".
When you want to mea-
sure once (such as for
long-term recording)
When measuring multi-
ple times (up to four
These are selected by setting memory partitioning.
See:
"Setting the Method for Recording to Internal Memory (Partition)" (p. 63)
Recording start and stop operations differ according to
whether memory partitioning is enabled.
• When memory partitioning is disabled
Only one-time measurement is available, but available
recording time is longer than when memory partitioning is
enabled.
• When memory partitioning is enabled
Four measurement sessions can be recorded, but only
one quarter of the amount of data can be recorded (com-
pared to when memory partitioning is disabled).
See:
"2.5 Internal Operating Status and Memory Usage" (p. 34)
"Appendix 1 Interval and Recording Time Settings" (p. A1)
Up to 50 events can be recorded, whether or not memory is partitioned. Verify the
number of recordable events before starting recording.
Summary of Contents for 3197
Page 1: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...
Page 2: ......
Page 8: ...Contents vi ...
Page 30: ...1 3 Measurement Flowchart 22 ...
Page 44: ...2 5 Internal Operating Status and Memory Usage 36 ...
Page 86: ...4 6 Initializing the Instrument System Reset 78 ...
Page 134: ...6 4 Viewing Recorded Data REVIEW State 126 ...
Page 176: ...9 4 Disposing of the Instrument 168 ...
Page 199: ......
Page 200: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...