![Hioki 3197 Instruction Manual Download Page 128](http://html.mh-extra.com/html/hioki/3197/3197_instruction-manual_2137178128.webp)
6.3 Viewing Anomalous Phenomena (EVENT screen)
120
The event waveform is displayed for the event selected in the Event List.
Pressing the up and down cursor keys sequentially changes the selected
event waveform.
「」
Displaying Detected Event Waveforms
[WAVEFORM]
Select the screen.
The Event Detail List can be scrolled up and
down when it contains more than four items.
The number of channels of displayed wave-
forms depends on the selected wiring configu-
ration.
Scale per division:
1 kV/div, or 500, 200, 100, 50, 20, 10 or 5 V/div
Vertical magnification can be selected from
factors of 1/2, 1, 2.5, 5, 10, 25, 50 and 100.
Available multiples are 1, 2, 4 or 8 times hor-
izontally.
Changing magnification of the
vertical axis
To change current magnification
Scale per division:
Available settings are multiples of the current
range.
Selects voltage
waveforms and
changes vertical
axis magnifica-
tion
Select an event.
Select voltage or current waveforms.
Display voltage waveforms.
Display current waveforms.
Available settings are multiples of
the current range.
Selects current waveforms
and changes vertical axis
magnification
Changes time axis
magnification
To change voltage magnification
To change time axis magnification
Summary of Contents for 3197
Page 1: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...
Page 2: ......
Page 8: ...Contents vi ...
Page 30: ...1 3 Measurement Flowchart 22 ...
Page 44: ...2 5 Internal Operating Status and Memory Usage 36 ...
Page 86: ...4 6 Initializing the Instrument System Reset 78 ...
Page 134: ...6 4 Viewing Recorded Data REVIEW State 126 ...
Page 176: ...9 4 Disposing of the Instrument 168 ...
Page 199: ......
Page 200: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...