89
Chapter 4 Analysis of Recorded Data
If an error occurs during data recording or file
operation, the error information is recorded on the
2nd line of recorded data that is saved on the PC.
This information can be checked when the file is
loaded to Excel or Memo Pad.
The following describes the character strings and
contents of error information.
Table of Error Indications
4.9 Error Information
1. Errors arising from the 3144-20
Error
Description
Stop By Battery
Low
Recording was stopped due to low battery
voltage. Replace the batteries of the 3144-20
with new ones.
Stop By Memory
Error
Recording was stopped due to frequent errors
that occurred in 3144-20 memory. The error is
related to flash memory of the 3144-20.
The 3144-20 must be repaired.
Stop By FIFO
Error
Recording was stopped due to a FIFO error
that occurred in 3144-20 memory. The error is
related to flash memory of the 3144-20.
The 3144-20 must be repaired.