Chapter 5 Specifications
100
5.2 Clamp Specifications
Sensor structure
Electrostatic coupling type of noncontact
voltage sensor
Frequency band
600 Hz to 30 MHz
(-3 dB or more with respect to 100 kHz)
Detecting sensitivity
Influence of conductor diameter
For Typical Characteristics, see Figure 1
(Condition: Conductor located in the cen-
ter of the clamp)
Fluctuation of sensitivity due to conductor
position
For Typical Characteristics, see Figure 2
Sensor output
Voltage output
Outputs voltage corresponding to the de-
tecting sensitivity.
Output impedance
50
Ω
±10%(1 kHz)
Max. clamp dia.
20 mm dia.
Maximum rated volt-
age to earth
200 VAC
Cord length
Approx. 1 m (39.37")
Functional ground
terminal
Provided
Supply voltage
±5 V (Supplied through connection to the
3144-20)
Maximum rated
power
0.5 VA max
Operating tempera-
ture and humidity
0 to 40°C (32 to 104°F), 80%RH or less
(non-condensating)
Storage tempera-
ture and humidity
-10 to 50°C (14 to 122°F), 80%RH or less
(non-condensating)
Operating environ-
ment
Indoors, altitude up to 2000 m (6562-ft.)