Alarm judgment and recording function
1.5.3 General Specifications
Recording mode
• Instantaneous value
• MAX/MIN/AVE
• Instantaneous value + MAX/MIN/AVE
Total 3 modes
Set the mode before the start of recording.
Recorded data
One data set contains time and temperature/humidity information (One channel each).
Data is scaled if scaling is ON.
Recording capacity
512 k bytes Flash memory
Quantity of recorded
data
• Instantaneous value recording mode: 30,000 data × 2 CH
• MAX/MIN/AVE recording mode: 15,000 data × 2 CH
• Instantaneous value + MAX/MIN/AVE recording mode: 12,000 data × 2 CH
Power outage
protection
After recovering from a power outage, the instrument automatically returns to the state
held before the outage.
Alarm judgment is made at every sampling, and the history will be recorded in flash memory if the measured
value recording function remains active.
Judgment method
Criterion threshold can be set to either Hi or Lo.
The instantaneous value at every sampling is judged (effective in any measurement
mode).
Recorded data
One data set contains time, generation/reversion, CH and judgment threshold.
Alarm output
Alarm output × 2 CH
Output is turned ON when an alarm (Hi or Lo) occurs.
Whether to hold the alarm output can be selected. The reset switch or a command
can be used to reset alarms.
Clock accuracy
±100 ppm (Reference value at temperature from 0 to 50°C (32 to 122°F) without the
communications module)
Backup
Recorded data (saved in flash memory)
Data loss for up to 2 minutes before and after a power outage may occur.
Communication
interface
CAN bus
Maximum rated
voltage to earth
33 Vrms, 70 VDC
Alarm output
Open collector: 30 VDC, 20 mA max.
Rated supply voltage 5 V±0.3 VDC
Maximum rated Power 1.4 W
Dielectric strength
1.4 kVAC Between input and alarm output, Input/Output and CAN bus (50/60 Hz,
Response current 5 mA, one minutes
Dimensions
Approx. 22.5W × 96H × 85D mm (0.89"W × 3.78"H × 3.35"D) (excluding projections)
Mass
Approx. 120 g (4.2 oz.)
Accessories
Ferrite clamp................................................................................................................ 1
Terminal block.............................................................................................................. 3
Option
9764-50 HUMIDIDTY SENSOR
Operational ranges
for temperature and
humidity
0 to 50°C (32 to 122°F), 80%RH or less (with no condensation)
Temperature and
humidity ranges for
storage
-10 to 50°C (14 to 122°F), 80%RH or less (with no condensation)
Operating
environment
Indoors, altitude up to 2000 m (6562-ft.)
Standards applying
Safety EN61010, Pollution degree 2, Measurement Category I
(anticipated transient overvoltage 330 V)
EMC
EN61326, CLASS A
Summary of Contents for 2300 Smart Site
Page 1: ...SMART SITE Instruction Manual 2300 February 2011 Revised edition 5 2300A981 05 11 02H...
Page 13: ...Contents XII...
Page 21: ...8 Service...
Page 87: ...74 4 2304 20 PULSE MODULE...
Page 105: ...92 5 2304 21 PULSE MODULE...
Page 253: ...240 12 2342 20 OUTPUT MODULE...
Page 275: ...262 14 2351 AIR MODULE 2352 20 WIRE MODULE...
Page 285: ...272 15 2353 20 LAN MODULE...
Page 317: ...304 18 2362 20 DC POWER MODULE...
Page 345: ...332 20 2392 01 02 MODULE BASE...
Page 347: ...334 21 Maintenance and Service...
Page 348: ......
Page 349: ......
Page 350: ......
Page 351: ......
Page 352: ......
Page 353: ......
Page 354: ......
Page 355: ......
Page 356: ......
Page 357: ......
Page 358: ......
Page 359: ......
Page 360: ......
Page 361: ......
Page 362: ......
Page 363: ......
Page 364: ......
Page 365: ......
Page 366: ......