5.4.3 Insulation of Internal Circuit
In the 2304-21, the input circuit and alarm output are insulated
from the CAN bus as shown in the block diagram below. (With-
stand voltage: 1.4 kVAC, 50/60 Hz, Response current: 5 mA, 1
minute)
The CH1 and CH2 terminals are not insulated from each other.
When measuring two measurement points having a potential differ-
ence, equalize the ground level potential of these two points, or use
another 2304-21 Module, since measurements may be adversely
affected.
Analog
circuit
CH1
/ CH2
Mea
s
ur
em
ent
Inp
u
t
Digital
circuit
In
sula
ting
cir
c
u
it
Module-to-
module com-
munications
circuit
CAN
bus
Con
nec
tor
Mod
u
le
bas
e
Pow
e
r
Insulation
withstand voltage
of 1.4 kVAC
Insulation
withstand voltage
of 1.4 kVAC
AL
ARM
A
la
rm
ou
tput
Ground
The COM terminal of the alarm output terminal is used for both
CH1 and CH2.
Summary of Contents for 2300 Smart Site
Page 1: ...SMART SITE Instruction Manual 2300 February 2011 Revised edition 5 2300A981 05 11 02H...
Page 13: ...Contents XII...
Page 21: ...8 Service...
Page 87: ...74 4 2304 20 PULSE MODULE...
Page 105: ...92 5 2304 21 PULSE MODULE...
Page 253: ...240 12 2342 20 OUTPUT MODULE...
Page 275: ...262 14 2351 AIR MODULE 2352 20 WIRE MODULE...
Page 285: ...272 15 2353 20 LAN MODULE...
Page 317: ...304 18 2362 20 DC POWER MODULE...
Page 345: ...332 20 2392 01 02 MODULE BASE...
Page 347: ...334 21 Maintenance and Service...
Page 348: ......
Page 349: ......
Page 350: ......
Page 351: ......
Page 352: ......
Page 353: ......
Page 354: ......
Page 355: ......
Page 356: ......
Page 357: ......
Page 358: ......
Page 359: ......
Page 360: ......
Page 361: ......
Page 362: ......
Page 363: ......
Page 364: ......
Page 365: ......
Page 366: ......