Commissioning
64/90
Step 1: Identify IP address of the device.
Ø
Open the
Ethernet Device Configuration
program.
Ø
Scan for devices.
Ê
Your analyzer device NANL-B500G-RE is displayed.
Figure 19: Ethernet Device Configuration - NANL-B500G-RE found (Example)
Ø
Read the IP address of the NANL-B500G-RE device whose firmware
you want to update. For the example shown in the figure above, the IP
address is
192.168.1.1
.
For details to the
Ethernet Device Configuration
program refer to
the
operating instruction manual
Ethernet Device Configuration,
Address setting for Ethernet capable Hilscher devices
.
Step 2: Upload current firmware image
Ø
Open a web browser.
Ø
Enter the IP address in the web browser's address bar and press the
Enter key.
Ê
The netANALYZER configuration page is displayed.
Ø
Below
Firmware image update
: Click
Browse
.
Ê
A dialog for file selection is displayed.
Ø
Select the firmware update file
nanl-b500g-re.update
. The
firmware update file is located on the product DVD in the folder
NANL-
B500G-RE firmware update
.
Ø
Click
Open
.
Ø
Click
Upload to device
.
Ê
During the upload process,
uploading
is displayed in front of the
Upload to device
. This process takes a few seconds.