187692(10) - EN - P280 Mattress Instructions for Use
Page 25
M
obi
liz
ing
the Pa
tie
n
t
Activating the CPR
Never allow an unqualified person to operate this function and ensure that no
obstacles (e.g., limbs, accessories, objects, power cords) or persons are
under the head section.
The CPR valve can be easily identified by the red
and yellow hose connector
on the left of the control
unit.
1. Disconnect the hose connector from the control unit.
The device takes approximately 15 seconds to deflate.
2. If necessary, lower the siderails and the head section of the bed, or place the
bed frame in CPR position before using the emergency flat return (refer to
the instructions of the bed manufacturer).
3. Place a CPR panel under the patient, or follow the CPR function protocols.
Canceling the CPR
1. Connect the hose connector to the control unit.
The device returns to the previous comfort settings.
2. If necessary, place the bed frame in the appropriate position (refer to the
instructions for the bed frame).