Standard Features
Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
35
5.
Repeat Step 3 and Step 4 on the other side of the bed.
NOTE:
After you have manually retracted the bed width, gently pull on each siderail to make sure that the slide
plates are locked in position.
To manually extend the bed
—do these steps
in this order:
CAUTION:
Caution
—Immediately after you manually extend the bed width, use the air supply unit to
inflate
the
side bolsters to fill in the gap between the siderails and the surface. See “Deflate and Inflate the Side
Bolsters and Foot Section” on page 65
1.
Pull the release lever below the
intermediate siderail
toward you as you pull
outward
on the
intermediate siderail as far as it will go.
2.
Pull the release lever below the
head siderail
toward you as you pull
outward
on the head siderail
as far as it will go.
3.
Repeat steps 1 and 2 on other side of bed.
NOTE:
After you have manually extended the bed width, gently push each siderail inward to make sure that
the slide plates are locked in position.
If power is restored to the bed, and you have not completed
all
width adjustments, these will occur:
•
Both indicators next to the
Retract
and
Extend
Width controls will flash if the bed is not in the
correct position and will continue to flash until the bed is
fully extended
or
fully retracted
.
•
A triple beep will sound every 10 seconds until the bed is adjusted to the
fully extended
or
fully retracted
position.
•
Err 6
will flash on the caregiver control pod until the bed is
fully
extended
or
fully retracted
.
•
If after 2 minutes the bed width has not been fully adjusted, a
continuous triple beep will sound until the bed reaches the
fully
extended
or
fully retracted
position.
•
Fully extend
or
fully retract
bed width to clear the
Err 6
code and
silence the triple beep.
•
Once the bed is fully extended or retracted, the indicator will come on solid and a single
confirmation beep will sound.
NOTES:
•
If the bed is not
fully retracted
or
fully extended
, you will not be able to raise the head of the
bed; you will only be able to lower it.
•
When power is restored, you can use the powered width controls to make the final adjustments.
FlexAfoot™ Bed Length Adjustment
CAUTION:
To help prevent equipment damage, obey these cautions:
•
Caution
—When you lengthen or shorten the foot end of the bed, make sure that the magnets
on the surface power cord do not interfere with the movement.
•
Caution
—Do not adjust the bed length when traction equipment is in use.