Service Calls
90
Compella™ Bariatric Bed System Instructions for Use (178951 REV 6)
T
HE
B
ED
CPR H
ANDLE
D
OES
N
OT
R
ETURN
TO
THE
D
ISENGAGED
P
OSITION
If the CPR handle does not fully return to the disengaged position, the head section may not rise and
hold the patient weight. To return the CPR handle to the disengaged position, make sure of these:
•
The CPR handle is free from bed linens and other equipment that could prevent its return.
•
The CPR handle is not damaged such that it drags excessively. To determine this, slightly push
the handle to see if it completely returns.
•
The CPR cable does not have a kink or other damage that may prevent the handle from fully
returning. You may need to contact your facility-authorized maintenance person or Hill-Rom
Technical Support to do this inspection.
SERVICE CALLS
WARNING:
Warning
—Only facility-authorized persons or Hill-Rom service technicians should service the bed.
Service done by unauthorized persons could cause injury and/or equipment damage.
When you call Hill-Rom about your unit, be prepared to give the serial number from the product
identification label. You will find the serial numbers in these locations:
•
Bed—as shown in the illustration
•
Air supply unit—on the patient-left side of the
unit.
•
Therapy surface—on the patient- right foot
corner of the cover
–
Topper—on the inside seam, on the patient-
right foot corner
•
Foam surface—on the bottom of the cover,
toward the foot end, in the center.
Bed Serial Number