WatchCare™ Incontinence Management System User and Service Manual (196414 REV 2)
37
1.3 Indicator Light Assembly Replacement
•
At the head end of the guide
tube, you may need to use a
screwdriver to move the
plastic cover enough to get
access to the old and new
cables.
•
After you route the new cable through the guide tube, do these:
a.
Pull the old cable out from the guide tube.
CAUTION:
Caution
—The foot section must be extended to correctly install the
indicator light. Otherwise, equipment damage could occur.
b.
Extend the foot section.
c.
Put the foot-end cover (B) in its location on the bed, and
install the four screws (A) to attach the foot-end cover (B) to
the bed (see Step 1 on page 36).
d.
Continue routing the new cable to the reader.
4.
Go to “Final Steps” on page 39.
Foot-End Cover
without
Holes
CAUTION:
When you set the sliding foot cover on the thigh section of the bed, be
careful not to accidentally disconnect the footboard sensor switch cable.
Equipment damage could occur.
1.
Remove the fourteen screws (C) that attach the sliding foot cover (D)
to the bed, and set the cover on the thigh section of the bed.
DRAFT 26-FEB-2019