18
Volara™ System, Model PVL1 User Manual (196286 REV C)
Assemble and Connect the Patient Circuit
WARNING:
Warning
—During the therapy, make sure the expiratory ports on the
clear spontaneous breathing adapter (G) are not blocked. This prevents
carbon dioxide from accumulating in the system.
6.
Connect and twist the parts until they
lock into position.
7.
Connect the appropriate patient
interface:
NOTE:
•
Mouthpiece
(K)—Insert and twist
the mouthpiece (K) into the
output port of the clear
spontaneous breathing adapter
(H).
•
Face mask
(L)—
a.
Insert and gently twist the
adapter (M) into the output
port of the clear spontaneous
breathing adapter (H).
b.
Connect the other end of the
adapter (N) to the face mask
(L).
•
Flexible tracheostomy adapter
(O)—
a.
Insert and gently twist the
adapter (M) into the output
port of the clear spontaneous
breathing adapter (H).
b.
Connect the other end of the
adapter (N) to the flexible tracheostomy adapter (O).
c.
Connect the other end of the flexible tracheostomy adapter
to the tracheostomy tube (patient’s own).
8.
Assemble the nebulizer kit and add the prescribed medication. See
page 22.