For detailed Warning and Caution statements, and operating 
instructions, refer to the 

VersaCare® Bed User Manual

 (USR119  

for A through J models; 161956 for K model and newer).

Control Pod Controls

Scale

Setup

1. Make sure the bed is plugged into AC 

power.

2. Put all linens, blankets, pillows, equipment, 

and standard items on the bed. 

3. Make sure none of the items on the bed 

touch the headboard.

4. Make sure the bed does not touch anything 

that could affect the patient weight (head-
wall, lines such as pendant controls, ventila-
tors, or drainage bags).

5. Make sure the patient is 

not

 on the bed.

6.

Zero the scale

 as follows:

NOTE

: The scale must be zeroed before the 

patient is put on the bed.

a.Press the 

Enable

 control.

b.Press and hold the 

Zero

 

control until 

00.0

 shows on the 

display (HOLD will show until 
00.0

 

shows).

NOTE

: After you release the Zero 

control, the display will flash CALC. Do not 
touch the bed until the display shows 0.0.

Weigh the Patient
NOTE

: For a more accurate reading, make sure 

the patient is lying still and is centered on the 
mattress.
1.

A through F models

, press the 

Enable

 

control. 

G model and newer

, go to step 2.

2. Press the 

Weigh

 con-

trol. The bed shows 
the current patient 
weight on the display. 

Change the Scale Units—Bed with the Bed 
Exit Alarm Silence Feature Only

The default units shown on the scale display is 
both pounds (lb) and kilograms (kg). To change 
the units to lb or kg only, do as follows: 
1. Make sure the Enable control indicator is off.

2. Press the 

Zero

 control for 

approximately five seconds. And 
then, as you continue to press 
the 

Zero

 control, press the 

Weigh

 control until you hear a 

beep. The display will be in con-
figuration mode with the current 
unit setting highlighted: lb only; 
kg only; or lb/kg.

3. Press and release the 

Weigh

 control to 

move through the settings. When you reach 
the applicable setting, release the control 
and wait until you hear a beep (approxi-
mately ten seconds). The display will store 
the new configuration and exit the configura-
tion mode.

NOTE: 

If you do not press the Weigh control 

within ten seconds, you will hear a beep to let 
you know that the selected configuration has 
been stored and the display is exiting configura-
tion mode.

Scale 

Bed

Summary of Contents for VersaCare

Page 1: ...VersaCare Bed Quick Reference Guide...

Page 2: ...ontrols 4 D Point of Care Caregiver Control Panel 5 6 E Control Pod 7 8 9 F IntelliDrive Transport System optional 10 G Active Integrated Response Treatment Surface or P500 Surface optional 12 H SafeV...

Page 3: ...either AC or battery for the Emergency Trendelenburg function to operate 1 Make sure the bed is plugged into AC power or the battery is activated 2 Pull and hold the CPR handle 3 Hold the CPR handle u...

Page 4: ...flash Battery Back up When the bed does not receive AC power bat tery back up can be activated When the bed operates on battery back up The functions on the caregiver control panel are available The...

Page 5: ...r each time you press another control Caregiver Control Panel Controls Chair 1 Set the brake 2 A through F models press the Enable control G model and newer go to step 3 3 Press and hold the Chair con...

Page 6: ...the bed to go into the 7 Tren delenburg position 2 Move the patient to the head of the bed as necessary 3 To deactivate the Boost function press any of the bed function controls except these Boost Sys...

Page 7: ...atient is lying still and is centered on the mattress 1 A through F models press the Enable control G model and newer go to step 2 2 Press the Weigh con trol The bed shows the current patient weight o...

Page 8: ...n The system has malfunctioned NOTE If a Bed Exit mode is active and the bed gets unplugged or the battery activated the Bed Exit Alarm System will be inactive When the bed is plugged in again the Bed...

Page 9: ...head angle alarm is active and the head section goes below the specified angle Bed Exit is in Silence mode bed with the Bed Exit Alarm Silence fea ture Alerts Flash Yellow and Green there is a techni...

Page 10: ...ed 7 Press one or both of the Enable switches on the transport handles 8 While you press the Enable switches push the transport handles forward or pull them backward for the applicable direction 9 To...

Page 11: ...er or the bat tery is active The Bed Down control is not locked out The Bed Not Down indicator is not flashing There is nothing between the upper frame and the base frame There is nothing between the...

Page 12: ...become very firm After 30 minutes if no other mode is selected the surface automatically returns to the Pressure Relief Normal mode NOTE If the surface is in Max Inflate mode and the bed gets unplugge...

Page 13: ...e siderails are not in the correct position or the head section is not below 25 an indicator will flash and an alarm will sound Deactivate 1 Press the Enable con trol 2 Press the opposite Turn control...

Page 14: ...solid To activate Sleep mode refer to Sleep Mode on page 13 The scale system is used correctly If it is not used correctly the surface may go into the non operational safe mode see All Surface Mode I...

Page 15: ...the sleep deck Use applicable precautions with gravity sensitive devices such as ventricular drains or lumbar drains before during and after you operate the bed functions Make sure flow rates on all g...

Page 16: ...ween 57 and 84 145 cm to 213 cm 2011 by Hill Rom Services Inc ALL RIGHTS RESERVED No part of this text shall be reproduced without written permission from Hill Rom Services Inc Active Integrated Respo...

Reviews: