4.1 Counterweight
Chapter 4: Removal, Replacement, and Adjustment Procedures
TotalCare® Bariatric Bed Service Manual (man299ra)
Page 4 - 7
4
Figure 4-1. Counterweight
Replacement
WARNING:
The counterweight weighs in excess of 75 lb (34 kg). Use two people
when removing or installing the counterweight. Failure to do so can
result in personal injury.
1. Install the counterweight (C) into the bed (D).
2. Install the three nuts (A) onto the screws (B) securing the counterweight
(C) into the bed (D).
3. Remove the Hi-Lo cylinder braces. Refer to “Hi-Lo Cylinder Brace—
TotalCare® Bed System Service Manual (man112)
4. Install the head end covers. Refer to “Cover—Head End” in the