Synclara™ Cough System, Model PSC1 User Manual (196286 REV E)
Synclara™ Therapy
Adjust the Patient Synchrony and End with Sigh settings
NOTE:
Note the number of treatment sessions carried out with the
connected patient circuit.
5.
Press
Back
to go to the automatic
therapy screen.
6.
To go to the
Home
menu, swipe the
Options
tab left and press the
Home
menu control.
NOTE:
Only available in Synclara™ Automatic mode.
Patient Synchrony
allows the system to synchronize cough cycles with
the patient’s breath pattern. When enabled, the therapy starts in
Pause
stage. When the patient’s inhale effort is detected, the
Inhale
stage
activates automatically.
The
End with Sigh
feature allows the system to help the patient open
up the collapsed alveoli during the last
Exhale
of the therapy.
1.
Select a preset therapy plan.
Number of times the patient circuit has been used
Maximum number of treatment sessions the patient
circuit can be used
Draft_E_2019-Apr-29_Cleaned