7.5 Auxiliary Power Supply—P442
Chapter 7: Accessories
RESIDENT® Low Bed Service Manual (man259)
Page 7 - 9
7
7.5
Auxiliary Power Supply—P442
Tools required:
None
The auxiliary power supply enables operation of the actuators to position the
bed in the event of a power failure. The auxiliary power supply charges
through a standard 120V AC, 60-cycle outlet, and can be fully charged in 24
hours. The auxiliary power supply allows up to 15 complete cycles of the
hilow and head section when fully charged.
Installation
1. Charge the auxiliary power supply (A) before a power failure (see figure 7-
Figure 7-2. Auxiliary Power Supply
2. During a power failure, plug the auxiliary power supply (A) into the
electrical control/lockout box.
Removal
Unplug the auxiliary power supply (A) from the electrical control/lockout box.
m259_053
Summary of Contents for RESIDENT P872
Page 2: ......
Page 6: ...Revisions Page iv RESIDENT Low Bed Service Manual man259 NOTES...
Page 12: ...Table of Contents Page x RESIDENT Low Bed Service Manual man259 NOTES...
Page 32: ...Chapter 2 Troubleshooting Procedures Page 2 2 RESIDENT Low Bed Service Manual man259 NOTES...
Page 78: ...Chapter 3 Theory of Operation Page 3 2 RESIDENT Low Bed Service Manual man259 NOTES...
Page 136: ...Chapter 5 Parts List Page 5 2 RESIDENT Low Bed Service Manual man259 NOTES...
Page 138: ...Warranty Chapter 5 Parts List Page 5 4 RESIDENT Low Bed Service Manual man259 NOTES...
Page 149: ...Main Frame Chapter 5 Parts List RESIDENT Low Bed Service Manual man259 Page 5 15 5 NOTES...
Page 158: ...Labels Chapter 5 Parts List Page 5 24 RESIDENT Low Bed Service Manual man259 NOTES...
Page 160: ...Chapter 6 General Procedures Page 6 2 RESIDENT Low Bed Service Manual man259 NOTES...
Page 170: ...Chapter 7 Accessories Page 7 2 RESIDENT Low Bed Service Manual man259 NOTES...