2.9 Automatic Contour Malfunction
Chapter 2: Troubleshooting Procedures
RESIDENT® Low Bed Service Manual (man259)
Page 2 - 45
2
Press and hold the head up button (A) on the hand pendant (see figure 2-18 on
page 2-39).
The head section rises, but the knee section remains flat.
Yes
No
↓
→
Replace the electrical control/lockout box (D) (refer to
procedure 4.9).