
2.8 Hilow Actuator Malfunction
Chapter 2: Troubleshooting Procedures
Page 2 - 38
RESIDENT® Low Bed Service Manual (man259)
Figure 2-17. Lockout Controls
9. Press and hold the hilow up button (A) on the hand pendant (see figure 2-
The bed stays still.
Yes
No
↓
→
If the bed moves, replace the electrical control/lockout box (refer
to procedure 4.9).
10. Press and hold the hilow down button (B) on the hand pendant.
The bed stays still.
Yes
No
↓
→
If the bed moves, replace the electrical control/lockout box (refer
to procedure 4.9).
11. Go to “Final Actions” on page 2-16.
m259_030
Summary of Contents for RESIDENT P872
Page 2: ......
Page 6: ...Revisions Page iv RESIDENT Low Bed Service Manual man259 NOTES...
Page 12: ...Table of Contents Page x RESIDENT Low Bed Service Manual man259 NOTES...
Page 32: ...Chapter 2 Troubleshooting Procedures Page 2 2 RESIDENT Low Bed Service Manual man259 NOTES...
Page 78: ...Chapter 3 Theory of Operation Page 3 2 RESIDENT Low Bed Service Manual man259 NOTES...
Page 136: ...Chapter 5 Parts List Page 5 2 RESIDENT Low Bed Service Manual man259 NOTES...
Page 138: ...Warranty Chapter 5 Parts List Page 5 4 RESIDENT Low Bed Service Manual man259 NOTES...
Page 149: ...Main Frame Chapter 5 Parts List RESIDENT Low Bed Service Manual man259 Page 5 15 5 NOTES...
Page 158: ...Labels Chapter 5 Parts List Page 5 24 RESIDENT Low Bed Service Manual man259 NOTES...
Page 160: ...Chapter 6 General Procedures Page 6 2 RESIDENT Low Bed Service Manual man259 NOTES...
Page 170: ...Chapter 7 Accessories Page 7 2 RESIDENT Low Bed Service Manual man259 NOTES...