2.9 Electric Knee Section Does Not Raise Or Lower Correctly
Chapter 2: Troubleshooting Procedures
Page 2 - 36
Resident
TM
LTC Bed Service Manual (man136rc)
2
Figure 2-27. Knee Controls
4. The knee motor actuator cable (A) is firmly connected to its receptacle on
the control box (on top of the hilow motor actuator), and the routing to the
control lockout shows no visible damage (see figure 2-28 on page 2-36).
Yes
No
↓
→
Plug the cable firmly into its receptacle, or replace the damaged
cable if necessary.
This solves the problem.
Yes
No
↓
→
Replace the knee motor actuator assembly (see “Knee Section
Motor Actuator” on page 4-20).
Figure 2-28. Knee Cable Routing Between the Control Box and the Control Lockout
A
B
B
A
m136_026
A
m136_027