2.1 No Functions Work
Chapter 2: Troubleshooting Procedures
Electrically Operated General Ward Bed
Page 2 - 7
Operation and Maintenance Manual
(154588 REV 1)
2
2.1
No Functions Work
This procedure assumes that no functions work either from the handset or the
Attendant Control Panel (ACP).
1. Verify that all functions are unlocked.
Yes
No
Unlock all functions. Go to “Function Checks” on page 2-3.
2. Verify that the cables for the handset, handset extension cable, ACP, and
control box are connected correctly.
Yes
No
Connect the functional cables. Go to “Function Checks” on page
2-3.
3. Verify that the mains cable is connected to an applicable power supply.
Yes
No
Connect the mains cable to the wall outlet, and proceed to
battery backup malfunction (if equipped). Go to RAP 2.6 on
page 2-15.
4. Unplug the mains power cable and use an applicable test device to check
for the presence of the voltage at the power outlet.
Yes
No
Inform the applicable individuals of the power outlet failure.
Move the bed to a power outlet that is functional.
5. Examine the mains power cable for damage.
Yes
No
Replace if damaged.
6. Contact Medicraft Hill-Rom Customer Service.