2.9 Brake or Steering Malfunction
Chapter 2: Troubleshooting Procedures
Page 2 - 18
Electrically Operated General Ward Bed
Operation and Maintenance Manual
(154588 REV 1)
2.9
Brake or Steering Malfunction
1. Make sure that the screws/pedals/linkages are not damaged or missing
from the castor braking system.
All hardware is present and in good condition.
Yes
No
Replace the hardware as needed.
2. The brakes are adjusted correctly.
Yes
No
Adjust the castors. See “Castor” on page 4-14.
3. Replace the castor(s). Does the problem continue?
Yes
No
Go to “Final Actions” on page 2-6.
4. Replace the backrest motor. Does the problem continue?
Yes
No
Go to “Final Actions” on page 2-6.
5. Contact Medicraft Hill-Rom Customer Service.