Bedienungsanleitung
Iontophorese-Therapiegeräte
HIDREX GS400
HIDREX PSP1000
www.hidrex.com
Gesundheit neu erleben
User Manual
Manua
H
I
DREX connect
ION
Page 1: ...Bedienungsanleitung Iontophorese Therapieger te HIDREX GS400 HIDREX PSP1000 www hidrex com Gesundheit neu erleben User Manual Manua HIDREX connectION...
Page 2: ...etup for Hands and Feet Treatment 13 Hints On Conducting Therapy 14 Conduct Therapy 15 Primary Settings And Saving Of Profiles 16 Begin Treatment 18 Important Advice for Conducting Treatments 20 Care...
Page 3: ...oid direct contact with the metallic surface i Several devices may not be simultaneously used by one patient i Prior to treatment remove any metallic jewelry wedding bands etc which would otherwise be...
Page 4: ...the two electrodes For example the current path for a person treating hands would be from one hand in the tray with an electrode up the arm across the chest down the other arm and hand to the other e...
Page 5: ...device and have it checked by an authorized HIDREX repair center i Do not perform iontophoresis therapy in parallel with aluminum containing antiperspirants in the same treatment area i Simultaneous...
Page 6: ...of the condition maintenance therapy involves one to three weekly sessions of approximately 15 minutes each Hint It is up to the doctor to decide whether the initial therapy should be partially or fu...
Page 7: ...and was treated with the anode and the left hand was treated with the cathode then on the next treatment the right hand would be treated with the cathode and the left hand would be treated with the an...
Page 8: ...lows you to change the pulse width in 10 increments for example ON 60 and OFF 40 or ON 70 and OFF 30 etc This allows for more effective treatment AND a better comfort level It is customizable for your...
Page 9: ...System Components and Scope of Delivery I Control Unit j Control unit main ON OFF switch main power switch k Connector for AC adapter 12V DC l Jacks for connecting the cable set to the treatment elec...
Page 10: ...Europe South America parts of Asia Type G UK i e Great Britain Malaysia Singapore Hong Kong Type I AUS i e Australia China Remove mains plug adapter by pressing the catch at the bottom side powerful...
Page 11: ...t electrodes As standard the electrodes are made of stainless steel but patients with a chrome nickel allergy also have the option of having electrodes made of aluminum VI Optional Accessories The opt...
Page 12: ...e the device in a well lit room on a firm level surface and make sure you can easily disconnect it from the power supply at any time Ensure a wall outlet is within reach and do not carry out the thera...
Page 13: ...e skin surfaces to be treated are well submerged The tops of the hand or feet should not be covered with water unless those areas need to be treated Simultaneously Treatment of hands and feet For the...
Page 14: ...be tolerated If the therapy does not have the expected effect on the feet you should treat the hands and feet separately Hints On Conducting Therapy Before starting treatment thoroughly remove residue...
Page 15: ...e off during the therapy In most cases this causes a safe but uncomfortable electric shock electric fence effect v Reduction of dosage before and during treatment and choice of current type DC PC or V...
Page 16: ...minutes on the left and the set dose in volts is shown flashing on the right 2 Select desired profile If a profile is selected one of the LED lights up the device will automatically suggest the respe...
Page 17: ...rofile or duration is disabled The device must be turned off briefly and then turned back on again to change the profile or duration 1 Select the dose within the 4 60V adjustment range according to yo...
Page 18: ...nt pulse width 100 thresholds for hands feet or armpits before you start the first treatment session To conduct axillary treatments as a rule the dose should not be raised above 15 volts because the s...
Page 19: ...e flashes again and shows the selected value With stopped therapy the treatment time and profile can no longer be changed The dose can then be changed with the or keys and the pulse with by holding ke...
Page 20: ...treatment the display will alternate showing the voltage and ot In this case you can continue the therapy without danger If the value exceeds the limit for a longer period of time the device first re...
Page 21: ...tance increases higher voltage values can be achieved if necessary Hint The lower the pulse width settings the more sensitive the protective circuit is and therefore a higher chance of dose limitation...
Page 22: ...ponge applicators 1 which can be re used to every session Between treatments the electrodes and trays hard case trays or ergonomic trays must always be cleaned dried and disinfected as described in th...
Page 23: ...build up can be removed with a common descaler or even vinegar or citric acid A discoloration of the electrode metal after the first couple of therapy sessions is normal does not affect efficacy and...
Page 24: ...mic tubs need to be disposed of and replaced by new ones The control unit and accessories hard case trays cables must be purified and disinfected A functional check and safety inspection must be compl...
Page 25: ...s in use In that case try non carbonated mineral or table water instead to raise the current Operational Check Proceed with the following steps for an operational check of your system 1 Set up the the...
Page 26: ...gal Requirements The regulations valid in the respective operator s country e g infection prevention technical servicing or registration and documentation regulations are to be kept in connection with...
Page 27: ...12 42579 Heiligenhaus Germany info hidrex de www hidrex com CE Mark ID of Notified Body Conformity with health and safety requirements set out in European Directive 93 42 EU UDI Device Identification...
Page 28: ...aware that adding accessories or components or modifying the powered device may take it more susceptible to interference from radio wave sources Note There is no easy way to evaluate their effect on...
Page 29: ...voltage dips flicker following IEC61000 3 3 Complaint Table 2 Electromagnetic Immunity Susceptibility IEC 60601 1 test level Actual level ESD IEC 61000 4 2 6kV cd 8kV ad 6kV cd 8kV ad Bursts IEC 6100...
Page 30: ...re from PE Polyethylene and outside package are of cardboard Dispose all package parts in an environmentally acceptable way If the device can no longer be used please dispose of it properly The nation...
Page 31: ...emperature 10 C to 30 C Rel humidity 30 to 70 Air pressure 700 hPa to 1060 hPa Output Treatment Voltage 4 60 VDC Max Current 35 mA Treatment Current 0 30 mA 5 mA reserved for the device Max output Pow...
Page 32: ...m2 Max Density at 30 mA mA cm2 Peak Density at 35 mA mA cm2 Hand Feet Hard Shell Case Individual Water Surface 818 0 037 0 043 Hand Feet Ergonomic tube individual Water Surface 595 0 05 0 059 Hand Fee...