Software description
N > ROTOR MAX
PROG
T/°C
RCF
RPM
t/min
1 20 9/;\9 1500 2:00
PROG
T/°C
RCF
RPM
t/min
RPM = 1500
PROG
T/°C
RCF
RPM
t/min
RPM = 800
PROG
T/°C
RCF
RPM
t/min
(1) 20 9/;\9 800 2:00
PROG
T/°C
RCF
RPM
t/min
Runtime (t/hms)
1 20 9/;\9 1500 2:00
PROG
T/°C
RCF
RPM
t/min
t/hms 0h < 1m> 0s
PROG
T/°C
RCF
RPM
t/min
t/hms 0h < 4m> 0s
PROG
T/°C
RCF
RPM
t/min
t/hms 0h 4m <20s>
PROG
T/°C
RCF
RPM
t/min
t/hms < 0h> 4m 20s
PROG
T/°C
RCF
RPM
t/min
(1) 20 9/;\9 1500 2:00
PROG
T/°C
RCF
RPM
t/min
If in doubt you should obtain clarification from the manufacturer.
If the speed in the activated program is higher than the
maximum rotor speed (Nmax), no centrifugation run can be
started. ‘N > ROTOR MAX’ is displayed
Dual centrifugation:
The maximum permissible speed of the applications
depends on the samples and sample containers. Every
sample container is approved only up to a certain rotational
speed depending on the design and application; this must
not be exceeded under any circumstances!
To set the speed (RPM), proceed as follows:
1.
Press the
[RPM] key.
2.
Use the
[adjusting knob] to set the desired value.
3.
To accept the settings, press the
[START] key.
4.
Save the program
Chapter 7.1.2 ‘Save programs (STO 1-99)’
7.2.5
Runtime (t/hms)
To set the continuous run, the minutes, seconds and hours
must be set to zero. The continuous run is indicated by the
@
symbol on the display. Dual centrifugation rotors must not
be operated in the continuous run.
To set the runtime, proceed as follows:
1.
Press the
[TIME] key.
🡆
The value shown in parentheses can be changed.
2.
Set the minutes with the
[adjusting knob].
3.
Press the
[TIME] key.
4.
Set the seconds with the
[adjusting knob].
5.
Press the
[TIME] key.
6.
Set the hours with the
[adjusting knob].
7.
To accept the settings, press the
[START] key.