en
12.2020 / Rev. 02
AB66005
126 / 174
Troubleshooting
1
[Logbook] button
2
[Close] button
8.2
Warnings and error messages
When a malfunction occurs, there is an optical and acoustic alarm
and the potential-free alarm contact switches.
Additional signal encoder for monitoring the con-
troller unit:
Alarms, which are sent from the operating unit or
controller unit, are placed at the potential-free alarm
output and can be connected to external signalling
equipment (acoustic or visual) by the customer.
In the event of alarms, the operating unit as well as
the controller unit outputs an acoustical signal.
This is adjustable on the operating unit. If the oper-
ating unit fails, only the controller unit beeps. If the
controller unit fails, the operating unit beeps.
The corresponding error code will appear on the
touchscreen. This additional beep also occurs in the
event of a door alarm.
The error can be queried via the
[Logbook] button. As long as the
error message is not actively closed, it remains in the upper section
of the screen.
The alarms are shown graphically in the run-time diagram, so that an
error message which was actively clicked away is still comprehen-
sible. In addition, the alarms and events can be called via
‘Settings
Logbook’
If the closing of the alarm is prevented by a PIN prompt, please con-
tact a service technician.
When you contact the Service department, please report the error
number that is displayed.
Fault description
Cause
Remedy
Program start not possible.
A program is already running
or a program start is planned.
Stop the running program.
The control contact is not
active when program opera-
tion has started, although it
was activated in the program
segment.
In the start conditions, the
control contact was not acti-
vated.
Activate the control contact when starting
the program
Date and time are not correctly
displayed.
Daylight savings time and
standard time must be cor-
rected manually.
Set the date and time in the device set-
tings
No.
Fault description
Cause
Remedy
2
Error 2
RAM memory is full.
Restart the system. Check whether
further operation is possible.
Summary of Contents for HettCube 200
Page 151: ...151 174 en 12 2020 Rev 02 AB66005 Index W Warning symbols 7 Warnings and error messages 126 ...
Page 152: ...en 12 2020 Rev 02 AB66005 152 174 Index ...
Page 153: ...13 Appendix ...
Page 154: ......
Page 166: ...en 12 2020 Rev 02 AB66005 166 174 Examples ...
Page 170: ...en 12 2020 Rev 02 AB66005 170 174 Accessory overview ...
Page 171: ...171 174 en 12 2020 Rev 02 AB66005 Declaration of conformity D Declaration of conformity ...