105 / 174
en
12.2020 / Rev. 02
AB66005
Operation
Fig. 164: ‘Language’ menu
Fig. 165: ‘Sound’ menu
1
[Buttons] controller
2
[Program end] controller
3 Input of
‘Number of reproduced
signal sounds’
NOTICE
−
If the ventilation opening is opened during cooling
mode, there is a danger of moisture from the envi-
ronment getting into the incubator.
−
At target temperatures below 10 °C, the ventila-
tion opening must be kept closed. Otherwise,
there is a danger the evaporator will ice up.
−
If the ventilation opening is open
−
the spatial temperature deviation will increase
by ± 0.1 °C as compared to the value speci-
fied in the technical data.
−
the current consumption specified in the tech-
nical data will increase, depending on the
ambient temperature, by approx. 0.016 kW/h.
−
In heating mode, when the ventilation opening is
open continuously, there is a risk of the biological
sample material drying out.
6.13.7
Language
In this menu, you can select the display language.
The display language is set during commissioning. If
another language is desired for the displays, these
can be selected from a list. The displayed language
changes directly after selecting a new language set-
ting.
To select the desired language, tap it.
6.13.8
Sound
The volume for the keypad feedback can be set in 4 levels or muted
with the
[Buttons] controller.
The volume can be set in 4 levels or switched off for the signal sound
after the end of the program via the
[Program end] controller.
Via the input
‘Number of reproduced signal sounds’, it can be speci-
fied how often a signal sound should repeat itself (1 to 10000 and
infinitely).
Summary of Contents for HettCube 200
Page 151: ...151 174 en 12 2020 Rev 02 AB66005 Index W Warning symbols 7 Warnings and error messages 126 ...
Page 152: ...en 12 2020 Rev 02 AB66005 152 174 Index ...
Page 153: ...13 Appendix ...
Page 154: ......
Page 166: ...en 12 2020 Rev 02 AB66005 166 174 Examples ...
Page 170: ...en 12 2020 Rev 02 AB66005 170 174 Accessory overview ...
Page 171: ...171 174 en 12 2020 Rev 02 AB66005 Declaration of conformity D Declaration of conformity ...