HERSILL OXYMO Quick Start Manual Download Page 1

QUICK START GUIDE                                                                     OXYMO 

 

1.  C

ONNECTION TO AN OXYGEN CYLINDER

 

 

Check the oxygen cylinder (1) is closed.  Connect  the 

pressure regulator (2)  (in the figure, Hersill P-300). 

Open the cylinder valve slowly. Check for leaks and that 

the cylinder has enough oxygen (the arrow points to the 

green area).  

Connect the regulator OXYMO P-15 MICRO (3) to the 

cylinder regulator.  

Do not connect the  OXYMO  device  to any other 

regulator than the OXYMO P-15 MICRO 

 

 

Connect the spiral hose to the regulator OXYMO P-15 

MICRO.  

 

 

Connect the other terminal of the spiral hose to the inlet 

connection of the unit. 

Insert batteries (back door of the device) and/or connect 

de AC/DC adapter.  Turned-on by pressing the yellow 

button. During the boot sequence the display will blink 

three times and the buzzer will beep  

Does not turn on OXYMO with nasal cannula connected 

to the patient 

 

After powering on, the display shows the last flow 

setting (in the figure, 3.5 L/min).  

   To set a different value: Press the b / - (

if the 

keyboard is blocked, the flow cannot be changed)

 

 

   To confirm the set: Press the button OK.. 

 

 

Connect the nasal cannula to the inlet connection of the 

device and to the patient.  

Check that  with  each breath,  a 

bolus

  of  oxygen  is 

released: you can see on the screen the symbol 

 

 

To properly  detect  inspiratory effort,  intranasal  tubes 

should not be too thin (HERSILL advised to be  flared-

tipped)  

 

To switch off the device: 

   Press the button ON / OFF. 

   Confirm by pressing OK.   

 

 

 