Lothian II

™ User/Service Manual 

 

 

 

 

 

 

 

 

 

 

 

 

 

Herida Lothian II 

 

 
 

 

 
 
 
 

NB: Please ensure this document is fully read & understood  

by the Clinician / Carer / Service Engineer prior to use. 

 
 

User & Full Service Manual 

Pressure Relieving / Alternating Air Mattress & Pump 

 

Pump Model: HH/P09  Mattress Model: HH/T06 

 

Summary of Contents for HH/P09

Page 1: ...ce Manual Herida Lothian II NB Please ensure this document is fully read understood by the Clinician Carer Service Engineer prior to use User Full Service Manual Pressure Relieving Alternating Air Mat...

Page 2: ...2 P a g e A U G U S T 2 0 1 9 Lothian II User Service Manual...

Page 3: ...tatic Mode Alternating Mode Profiled Bed Clinical Considerations Installation Safety Instructions Pump Operation Mattress Pump Characteristics MAINTENANCE 10 Cleaning Maintenance Repair Service Interv...

Page 4: ...ight Do not use phenolic based cleaning solutions Keep sharp objects away from the mattress Ensure that the system is clean and dry prior to storage Store the pump and mattress in the protective bags...

Page 5: ...physically ventilates thus allowing immediate relief between cycles rather than awaiting deflation under patient weight as is often the case Application This product is to be used for those patients...

Page 6: ...for 5 minutes This is one complete cycle Profiled Bed Mattress Clinical Carer Considerations If the bed is in the profiled position the patient is touching the foam base ensure that the pump pressure...

Page 7: ...that there are no sharp or hard objects on the bed before unfolding the mattress laying out flat The air inlet should be located at the feet side of the bed 3 Use the Rapid Inflator if available to i...

Page 8: ...kes place the digital display will show The pump will revert back to the previous alternating cycle after 30 minutes Pressure Monitor The light is illuminated when pressure is normal extinguished when...

Page 9: ...ineer assistance to rectify Power Off Alarm The alarm will sound if power is lost to the pump Pump Characteristics Power AC220V AC110V Pressure Range 10 52 mmHg Internal Fuse F0 5A 250V Plug Fuse 3A F...

Page 10: ...have been trained by Herida Healthcare Ltd and only original manufacturers parts should be used Failure to do so could result in a void warranty situation Service Interval Instructions This product m...

Page 11: ...is should be done via a pressurised system soak test over at least 3 cycles or under weights to replicate patient in situ The tubing and connectors that run down the side of the mattress must be clear...

Page 12: ...l fuse is F0 5A 250V When replacing the fuse ensure a similar type is used and that the fuse is disposed of in the appropriate manner During normal working when the emergency deflation is needed switc...

Page 13: ...luding cells tubing and CPR toggle Ensure all cell fasteners are snapped to the mattress base and are not lose or faulty Check that stitching which secures the straps to the base of the mattress is so...

Page 14: ...14 P a g e A U G U S T 2 0 1 9 Lothian II User Service Manual P09 Pump...

Page 15: ...against electric shock Class II electrical appliance Double insulated Dispose of electrical equipment correctly WEEE Do not use broken or damaged electrical plugs or cables as this may lead to electri...

Page 16: ...nnectors Repair the air tubes or call for replacement Re attach CPR plug 3 Mattress is too hard or too soft The air pressure is not adjusting to required level Adjust the pump pressure according to th...

Page 17: ...1HE Freephone 44 800 193 6030 Email sales heridahealthcare co uk Company Number 9918019 VAT Registration Number 233587691 Disclaimer Please be aware that all products should be disposed of in an envi...