PowerSpin

  LX Centrifuge 

with Variable Speed and Timer 
 
Item # 112-6907 with 6 place rotor 
Item # 112-6909 with 6 place rotor 
 
 
 
OPERATION MANUAL 

 

 
 
 
Revision Date 2014/03 
 
 
 
 
 
 
 

 

Distributed by (in US only):

 

 

(manufacturer) 

HENRY SCHEIN INC.  

135 Duryea Road              
Melville, NY 11747 USA 

 

 

Henry Schein U.K. Holdings Ltd. 

Medcare House, Centurion Close 
Gillingham Business Park 
Gillingham ME8 0SB U.K. 

 

 

 

 

 

Summary of Contents for PowerSpin 112-6907

Page 1: ...e rotor Item 112 6909 with 6 place rotor OPERATION MANUAL Revision Date 2014 03 Distributed by in US only manufacturer HENRY SCHEIN INC 135 Duryea Road Melville NY 11747 USA Henry Schein U K Holdings Ltd Medcare House Centurion Close Gillingham Business Park Gillingham ME8 0SB U K ...

Page 2: ...es Can spin 2 ml and 4 5 ml tubes with available adapters Linear speed control allows use from 500 rpm 4 000 rpm Dependable mechanical 30 minute timer with bell signal at end of run High impact plastic housing for years of dependable use Safety switch in lid shuts off motor when lid is unlatched Latch release located on the right front side of unit Power indicator light tells when unit is powered ...

Page 3: ...ing styrofoam disc is packed inside the lid of the centrifuge to prevent the rotor or shields from moving during shipping REMOVE THIS PACKING DISC PRIOR TO OPERATION OF THE CENTRIFUGE The centrifuge rotor is pre installed No rotor installation and calibration is required The power inlet is located in the back of the centrifuge Plug in one end of the power cord to the inlet and the other end to a p...

Page 4: ...ng is not necessary A surface cleaner such as Formula 409 can be used to clean the outside of the instrument The rotor chamber and cover can be disinfected with a mild solution of bleach or other disinfecting liquid and allowed to dry Do not drench or immerse the unit in any liquid Be sure to disconnect the power before attempting to clean Routine Rotor Inspection and Maintenance All centrifuge ro...

Page 5: ...ge Do not operate with a damaged cord Biohazard If a tube breaks a biological hazard may exist Unplug the instrument Clean using approved laboratory procedures Always wear the proper protective clothing and gloves Use a suitable disinfectant and dispose of the sample properly Refer to the guidelines set forth by OSHA CLIA etc Service If you feel that your centrifuge needs factory service please ca...

Page 6: ... turned on b Lamp needs to be replaced c Fuse is blown 4 Fuse Replacement CAUTION DISCONNECT THE POWER CORD BEFORE REPLACING FUSE Follow the procedures below to replace the fuse a Unplug the power cord Turn the on off switch to off b Remove the power cord from the power inlet on the back of the centrifuge c Locate the fuse holder The fuse holder is a part of the power inlet On off switch Inlet Fus...

Page 7: ...__________500 4 000 rpm Speed accuracy 100 rpm 3000 RPM Width________________________________10 2 inches Depth________________________________11 8 inches Height________________________________8 inches Weight________________________________18 lbs ...