Helmer Scientific Automatic Cell Washing System - UltraCW II
Operation Manual
360386/D
20
Error Code
Error Name
Description
Possible Causes
Miscellaneous
50
Unknown
An error has occurred, but the kind of the error
could not be identified
•
Unexpected behavior
51
Program Interrupted
A running program was interrupted.
•
Power interrupted during a running program
52
Program Aborted by User The program was aborted by the user
•
User has aborted the program
53
Imbalance
The program was stopped because of an
imbalance of the rotor
•
Rotor was not loaded symmetrically
•
Positioning of the imbalance sensor not correct