Helmer Scientific i.Series
®
- Platelet Agitator
Operation Manual
360390/A
11
4 Specifications
4.1
Operating Standards
These units are designed to operate under the following environmental conditions:
♦
Indoor use only
♦
Altitude (maximum): 2000 m
♦
Ambient temperature range: 15 °C to 35 °C
♦
Relative humidity (maximum for ambient temperature): 80% for temperatures up to 31 °C, decreasing linearly to 45% at 35 °C
♦
Overvoltage category: I
♦
Pollution degree: 2
♦
Sound level is less than 60 dB(A).
♦
Mains supply voltage: ±10% of nominal voltage
♦
RF Emissions: Group 1 - Class A
♦
EMC Environment: Basic
Note
Power draw is measured in full-load Amperes during stand-alone operation and includes power supply.
Table 2. Electrical Specifications
PF15-Pro
PF48-Pro
PF96-Pro
Input Voltage and
Frequency
Stand-alone (power supply): 100-240 VAC, 50/60 Hz
Agitator Unit: 24VDC
Voltage Tolerance
±10% (AC input to power supply)
Power Draw
< 16 Watts nominal
Power Source
24VDC AC/DC power supply (stand-alone)
24VDC umbilical cable (configured with incubator)
Agitation Speed
(cycles / minute)
40-80 CPM
Remote Alarm Capacity
1 A at 33 V (AC) RMS or 70 V (DC)
Table 3. Agitator Specifications
PF15-Pro
PF48-Pro
PF96-Pro
Height
13.5” (343 mm)
14.9” (379 mm)
14.9” (379 mm)
Width
16.0” (407 mm)
17.8” (453 mm)
32.8” (834 mm)
Depth
9.1” (232 mm)
14.9” (379 mm)
14.9” (379 mm)
Weight
33 lbs (15 kg)
50 lbs (23 kg)
80 lbs (37 kg)
CAUTION
• The interface on the remote alarm monitoring system is intended for connection to the end user’s central alarm
system(s) that uses normally-open or normally-closed dry contacts.
• If an external power supply exceeding 30 V (RMS) or 70 V (DC) is connected to the remote alarm monitoring system’s
circuit, the remote alarm will not function properly; may be damaged; or may result in injury to the user.