Helios PD2-6400 High Voltage & High Current Meter
Instruction Manual
44
Meter
Operation
The meter is capable of accepting two input channels
(A and V) of either current (0-5 AAC/DC) or voltage
signals (0-300 VAC/DC) and displaying these signals
in engineering units from -99999 to 999999.
A power function channel (P) is available to perform the
power operation (P = IV) on channel A and V, with
adder and factor constants, and display the results.
Engineering units or tags may be displayed with these
three channels.
The dual-line display can be customized by the user.
Typically, display line 1 is used to display the channel
V, while line 2 is used to alternate between displaying
input channels A and P.
Additionally, the meter can be set up to display any
channel on line 1 and a unit or tag on line 2. The relays
and analog output can be programmed to operate
based on any channel.
Button Operation
Button
Symbol
Description
Press to enter or exit Programming
Mode, view settings, or exit max/min
readings
Press to reset max/min readings or
other parameter/function assigned
through the
User
menu
Press to display max/min readings or
other parameter/function assigned
through the
User
menu
Press to acknowledge relays or other
parameters/function assigned through
the
User
menu
Function Keys Operation
During operation, the programmable function keys
operate according to the way they have been
programmed in the
Advanced Features
– User
menu.
Programmable Function Keys User Menu
The table above shows the factory default settings for
F1, F2, and F3.
Digital Inputs Operation
Five (5) digital inputs, F4, DI-1 to DI-4, come standard
on the meter. These digital inputs are programmed
identically to function keys F1, F2, and F3. The inputs
are triggered with a contact closure to +5 in the case
of digital inputs 1-4 or with an active high signal, see
for details. The F4
is triggered with a contact closure to COM or with an
active low signal. During operation, digital inputs
operate according to the way they are programmed in
the Advanced Features
– User menu. See
Programmable Function Keys User Menu
) on
Maximum/Minimum Readings
The max & min readings (peak & valley) reached by
the process can be displayed either continuously or
momentary:
1. Display briefly by assigning to the F1-F3
function keys or to the digital inputs in the
User
menu.
2. Display continuously by assigning either
display to max/min through the
Display
menu.
Any of the F1-F3 function keys (buttons) and the
digital inputs can be programmed to reset the max &
min readings. The meters are set at the factory to
display the max reading by pressing the Up arrow/F2
button and to use the Right arrow/F1 button to access
the
Reset
menu.
To display max and min channel A reading using
function key with factory defaults:
1. Press Up arrow/F2 button to display minimum
reading of channel A since the last reset/power-
up. The display will then display the maximum
reading of channel A since the last reset/power-
up.
2. Press the Up arrow/F2 button again to display the
minimum reading of channel A since the last
reset/power up.
3. To reset max/min press Right arrow/F1 button to
access the Reset menu. The max & min displays
are reset to actual values.
4. Press Menu to exit max/min display reading.
MENU
MENU
RIGHT
RESET
F1
UP
MAX
F2
ACK
ENTER
F3