Assembly
Assembly
34
35
Tubes: pump head SP vario
Additional components:
1x Aid pin
Aid pin
Pump head SP vario allows operating
tubes of different wall thickness.
✓
Remove protection shield.
➜
Remove clamping screw to remove
clamp.
Clamping screw
Rotor
Clamp
Protection shield
➜
Turn vario wheel via aid pin to
maximum wall thickness.
➜
Mount tube in a curve of 180° into
tube bed.
➜
Turn rotor by hand till tube lies
between roller and tread.
➜
Turn vario wheel via aid pin to
minimum wall thickness till you
sense a noticeable resistance.
➜
Tighten fixiing screw to hold vario
wheelin this position.
Vario
wheel
Rotor
Fixing
screw
Aid pin
maximum
minimum
➜
Turn rotor repeatedly to ensure tube is crushed by rollers in whole width.
➜
Attach clamp with clamping screws.
➜
Mount protection shield and secure with knurled screws.
➜
Check correct mounting by running a short test.
The right pressure is reached if fluid is dispensed during operation but
stops passing through when pump is stopped.
Small tubes, inner diameter 0.8 mm and 1.7 mm
Tubes with an inner diameter of
0.8 mm and 1.7 mm are delivered
with 2 additional tube cuttings.
Push both tube cuttings onto the
dispenser tube.
➜
Place tube in pump head with cuttings
positioned in clamping sleeve.
✓
The cuttings ensure save clamping of
the small tube.
Aid pin Tube inner Ø 0.8 mm / 1.7mm
Tube cuttings
Tubes: pump head SP quick
Pump head SP quick features a quick-clamp
mechanic for clamping tubes.
➜
Move clamping lever to the left to open
tube bed.
➜
Turn both clamping screws counter-
clockwise as far as they will go for
maximum tube size.
Clamping screws
Clamping lever
Tube bed open
➜
Place tube into open tube bed.
➜
Move clamping lever cautiously to the
right to close tube bed.
✓
When sensing significant resistance
during closing either clamping screws
are not open or tube is too large.
➜
Turn both clamping screws clockwise to
clamp tube.
➜
Check correct mounting by running a
short test.
Tube bed closed
Summary of Contents for Hei-FLOW Advantage 01