85
Attachements
Pump head SP standard
To mount tubes on the SP standard pump head, proceed as follows:
➜
Remove the protector from the pump
head.
➜
Remove the clamping lever.
➜
Insert the tube into the pump head.
➜
Turn the pump head rotor in such a
way that the inserted tube is prefixed
between the rotor and the housing
(see illustration, several turns may be
necessary!).
➜
Mount the clamping lever.
➜
Slightly tighten the inserted tube: Pull
on both ends with moderate force.
➜
Pre-tensioning increases the service
life of the tube and ensures maximum
operational reliability.
➜
Mount the protector.
➜
Perform a short test run before
commissioning to ensure continuity of
the mounted tube.
Summary of Contents for 036150010
Page 2: ...Originalbetriebsanleitung Seite 6 48 Translation of the original instructions Page 56 103...
Page 13: ...13 Ger tebeschreibung Ger tebeschreibung bersicht Hei FLOW Core bersicht Hei FLOW Expert...
Page 14: ...14 Ger tebeschreibung bersicht Hei FLOW Ultimate Bedienfeld Hei FLOW Ultimate...
Page 32: ...32 Anhang Mehrkanalpumpenk pfe Lieferumfang...
Page 44: ...44 Leistungsbereich Anhang...
Page 49: ...Translation of the original instructions Page 56 103...
Page 53: ......
Page 61: ...63 Device description Overview Hei FLOW Core Overview Hei FLOW Expert Device description...
Page 62: ...64 Overview Hei FLOW Ultimate Control panel Hei FLOW Ultimate Device description...
Page 80: ...82 Attachements Multi channel pump heads Scope of delivery...
Page 87: ...89 Attachements Figure cassettes for multi channel pump heads...
Page 92: ...94 Performance range Attachements...
Page 99: ...101 RoHS Konformit tserkl rung RoHS Declaration of Conformity Zertifikate Certifications...