75
Attachements
Pin assignment analogue interface
PIN 1
+5 V
PIN 2
Start/Stop
PIN 3
0 – 10 V
PIN 4
Direction of
rotation
PIN 5
0 V (GND)
PIN 6
4 – 20 mA
Control voltage
Control current
Hei-FLOW Expert/
Ultimate 01
Hei-FLOW Expert/
Ultimate 06
0 – 0.9 V
0 – 4.9 mA
0 rpm
0 rpm
1 V
5 mA
5 rpm
24 rpm
10 V
20 mA
120 rpm
600 rpm
➜
TTL level 0 V = clockwise rotation, TTL level 5 V = counter-
clockwise rotation.
➜
TTL-pulse 5 V switches the rotation on and off again at the next
pulse. This function is always activated.
➜
For controlling the [direction of rotation] and [rotation start/
stop] functions, a voltage of 5 V is also available at the
interface socket in order to be able to switch with potential-free
contacts.
➜
The analogue interface is detected at a voltage of 0.2 V or a
current of 4 mA;
➜
Hei-FLOW Ultimate: When the interface is detected, the display
shows
[EA]
. As long as the pump drive is controlled via the
analogue interface, the display and all buttons except [Start/
Stop] are inactive.
Summary of Contents for 036150010
Page 2: ...Originalbetriebsanleitung Seite 6 48 Translation of the original instructions Page 56 103...
Page 13: ...13 Ger tebeschreibung Ger tebeschreibung bersicht Hei FLOW Core bersicht Hei FLOW Expert...
Page 14: ...14 Ger tebeschreibung bersicht Hei FLOW Ultimate Bedienfeld Hei FLOW Ultimate...
Page 32: ...32 Anhang Mehrkanalpumpenk pfe Lieferumfang...
Page 44: ...44 Leistungsbereich Anhang...
Page 49: ...Translation of the original instructions Page 56 103...
Page 53: ......
Page 61: ...63 Device description Overview Hei FLOW Core Overview Hei FLOW Expert Device description...
Page 62: ...64 Overview Hei FLOW Ultimate Control panel Hei FLOW Ultimate Device description...
Page 80: ...82 Attachements Multi channel pump heads Scope of delivery...
Page 87: ...89 Attachements Figure cassettes for multi channel pump heads...
Page 92: ...94 Performance range Attachements...
Page 99: ...101 RoHS Konformit tserkl rung RoHS Declaration of Conformity Zertifikate Certifications...