Information for OEM and Setup users | Servicing and maintenance
II
8.6.2
Diagnostics for encoders with EnDat interface
You can check the encoders with EnDat interface for proper functioning by reading
out the errors or warnings and by evaluating their functional reserves.
Depending on the encoder, not all of the functional reserves and messages are
supported.
Functional reserves
1
2
3
4
5
Figure 27: Example of functional reserves of a length gauge
1
Minimum percentage value of functional reserves at this specific position
2
Absolute track
3
Incremental track
4
Position value calculation
5
Current position of encoder
Path:
Settings
►
Axes
►
<Axis name>
►
Encoder
►
Diagnosis
Functional reserves
Parameter
Explanation
Absolute track
Displays the functional reserves of the absolute
track
Incremental track
Displays the functional reserves of the incremental
track
Position value calculation
Displays the functional reserves of position value
calculation
Position
Displays the actual current position of the encoder
The product displays the functional reserves in a bar graph:
Color
Range
Evaluation
Yellow
0 % ... 25 %
Servicing/maintenance recommended;
testing recommended (e.g., with PWT 101)
Green
25 % ... 100%
The encoder is within the specification
178
HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 06/2022
Summary of Contents for GAGE-CHEK 2000
Page 1: ...GAGE CHEK 2000 Operating Instructions Evaluation Unit English en 06 2022...
Page 3: ...I General information...
Page 7: ...1 Fundamentals...
Page 14: ...2 Safety...
Page 19: ...3 Basic operation...
Page 43: ...II Information for OEM and Setup users...
Page 48: ...1 Transport and storage...
Page 54: ...2 Mounting...
Page 60: ...3 Installation...
Page 72: ...4 Commissioning...
Page 121: ...5 Setup...
Page 150: ...6 File management...
Page 157: ...7 Settings...
Page 172: ...8 Servicing and maintenance...
Page 184: ...9 Removal and disposal...
Page 186: ...10 Specifications...
Page 194: ......
Page 195: ...III Information for the Operator user...
Page 240: ...14 What to do if...