Instruction Manual

Wrist Type 

Blood Pressure Monitor

Model BPM280XL

This manual and product are not meant to be a substitute for advice provided by 

your doctor or other medical professional. You are not to use the information 

contained herein, or this product for diagnosing or treating a health problem or 

prescribing any medication. If you have or suspect that you have a medical 

problem, promptly consult your healthcare provider.

The device is designed for non-invasive measurements and monitoring of adults’ 

arterial blood pressure. It measures human's systolic blood pressure, diastolic blood 

pressure and heart rate on wrist by oscillometric measurement method.

The device is recommended for use by people over the age of 18 and is for home 

use and not clinical use.

Blood Pressure Monitor Intended Use

*Attention!

To operate the device properly and get the best result from it, please read the 

Instruction Manual thoroughly before you begin measuring and keep it carefully 

so that you can access and consult it at any time.

Medical Disclaimer

1-1. What is blood pressure ?

1. About Blood Pressure

Having one's blood pressure measured by a doctor in a hospital or a clinic, and a 

group health checks, tends to stimulate nervousness in the subject and may even 

create high blood pressure, blood pressure also varies in accordance with a variety of 

conditions and so judgment is not possible on the basis of a single measurement. 

The blood pressure measured first thing in the morning after getting up, before taking 

any food and with the subject still, is known as the fundamental blood pressure. In 

practice it is rather difficult to record the fundamental blood pressure, but to come as 

near as possible to measuring the blood pressure in an environment that is close to 

this is why it is useful to take the measurement at home.

Blood pressure is a measurement of the force of blood flowing against the walls of 

the arteries. Arterial blood pressure is constantly changing during the course of the 

cardiac cycle. The highest pressure in the cycle is called the systolic blood pressure 

and the lowest is the diastolic blood pressure. Both of pressure readings, the systolic 

and the diastolic, are necessary to enable a physician to evaluate the status of a patient's 

blood pressure. Many factors such as physical activity, anxiety or the time of day can 

influence your blood pressure. Blood pressure is typically low in the morning and increases 

from the afternoon to the evening. It is lower in the summer and higher in the winter.

1-2. Why is it useful to measure blood pressure at home?

A. WHO blood pressure classifications

Standards for assessment of high or low 

blood pressure without regard to age, have 

been established by the World Health 

Organization (WHO), as shown in the chart.

Optimal

Normal

High normal

Mild hypertension

Moderate hypertension

Severe hypertension

120

130

85
80

95

100

90

105

110

140

150

160

170

180

* Figure No.1

B. Variations in blood pressure

* Figure No.2

Individual blood pressures vary greatly 

both on a daily and a seasonal basis. 

These variations are even more pronounced in hypertense patients. 

(Hypertense: means a person who has high blood pressure symptom.) 

Normally the blood pressure rises while at work 

and is at its lowest during sleeping period. The 

graph below illustrated the variations in blood 

pressure over a whole day with measurement 

taken every five minutes. (The thick line 

represents sleep. The rise in blood pressure at 4 

PM <A in the graph> and 12 PM <B in the 

graph> correspond to an attack of pain.)

 Please rest for at least 5 ~ 10 minutes before taking the measurement.

 To allow your blood vessels to return to the condition prior to taking the measurement, please wait at least 3 ~ 5 minutes 

   between measurements. You may need to adjust the wait time according to your personal physiological situation.

 If you have one of the circulatory problems as arteriosclerosis, diabetes, liver disease, kidney disease, severe 

   hypertension, peripheral circulation......, please consult your doctor or healthcare professional before using the devices.

 Blood pressure measurements should be interpreted by a physician or trained health professional who is familiar with 

   one's medical history. By using the unit regularly and recording the results for the physician to interpret, one can keep 

   physician informed of the continuing trends in one's blood pressure.

 Wind the cuff snugly around your wrist and the cuff must be at the same level as your heart.

 Do not move the unit during measurement, or the proper measurement will not be achieved.

 Take measurement in a relaxed position without talking.

 Do not wind the cuff over jacket or sweater sleeve, or measurement cannot be done.

 Keep in mind that blood pressure naturally varies from time to time throughout the day and also is affected by lots of 

   different factors such as smoking, alcohol consumption, medication and physical activity.

 People with a condition that causes circulatory problems (diabetes, kidney disease, arteriosclerosis or poor peripheral 

   circulation) may get lower readings with this monitor than with a blood pressure monitor that is used on the upper arm. 

   Please consult your physician to determine if your wrist blood pressure accurately reflects your actual blood pressure.

 Blood pressure measurements taken with the unit are equivalent to those obtained by a trained observer using the 

   cuff / stethoscope auscultation method and are within the accuracy limits prescribed by the American National Standard 

   for Electronic or Automated Sphygmomanometers.

2. Important information before use of the unit

 Wait 30 ~ 45 minutes before measurement if you have just consumed caffeinated beverages or smoked cigarettes.

 We recommend you to use the same wrist (preferably the left wrist) and to measure around the same time each day.

4-2. Irregular Heartbeat Detector (IHB)

*Note

Μ

 Any irregular heartbeat detected will not be stored in the memory.

The appearance of this symbol,              , signifies that a certain pulse irregularity was detected 

during the measurement. Talking, moving, shaking or an irregular pulse during the measurement 

can result in the appearance of this icon. Usually this is not a cause for concern, however if the 

symbol appears often, we recommend you seek medical advice. Please note that the device will 

not replace a true cardiac examination, but serves to detect pulse irregularities at an early stage. 

As a safeguard, we recommend that if you have arrhythmias such as atrial or ventricular premature 

beats and atrial fibrillation or any other special conditions you should check with your physician 

before using your BPM.

The hearbeat rhythm that is more than or less than 25% from the average heart beat intervals is 

usually defined as an irregular heartbeat rhythm.In order to filter the unstable status of user and 

avoid affecting the detection of heart rate from any movement, shaking or talking in the begining of 

measurement, the method of averaging heart beat intervals of subject device is calculated with the 

three proper heart beat pulses detected in the begining of measurement and that is different from 

a strict mathematical averaging of all recorded intervals.At least 3 beats with at least 25% difference 

from the average heart beat interval will generate the IHB symbol on the LCD.

*Attention!

 Batteries are hazardous waste. Do not dispose them together with the household garbage. 

   Please take the used batteries to the recycling collection point according to your local regulations.

 Keep the battery away from small children.

 To prolong the battery life and prevent damage caused by leakage, remove the batteries from the 

   device if the device is not to be used for a long period.

 Memories (if any) will not be deleted during battery replacement.

 After replacing the batteries, reset the date and time.

LCD DISPLAY

START/STOP

 BUTTON

RECALL MEMORY

BUTTON

WRIST CUFF

CARRYING CASE

ALKALINE
BATTERIES
UM-4 (AAA) 1.5V x 2

* Figure No.3

3-1. Name of the parts

3. Device Description

UP & DOWN BUTTON

FUNCTION

 BUTTON

3-2. Description of display symbols

* Figure No.4

4. Main feature: 

4-1. Personalized Target Limit Function

The device allows you to pre-set your personalized values of blood pressure which should be 

recommended or decided by your physician or medical professional. If the Systolic or Diastolic 

pressure of the measurement is greater than your pre-set personalized level, the digit of the result 

on the screen will flash to alert. If the Systolic or Diastolic pressure of the measurement is equal to or 

less than your pre-set personalized levels, or if you did not set the personalized levels, the SMILE 

FACE symbol will appear and flash on screen after each measurement.

4-3. 3D Positioning Technology

Keeping wrist at heart level while measuring is extremely important to get accurate blood pressure 

values. The device features a Wrist Position Sensor which guides you to adjust your hand by the 

WRIST DIRECTION INDICATOR symbols shown on screen and helps to determine the optimal 

height of the wrist for measurement.

HOUR: MINUTE

SYSTOLIC PRESSURE

IRREGULAR HEART BEAT SYMBOL

LOW BATTERY SYMBOL

DIASTOLIC PRESSURE

HEART RATE

PULSE SYMBOL

BLOOD PRESSURE UNIT

SELF-DIAGNOSTIC 

SMILE FACE

MONTH/ DATE

ERROR SYMBOL

MEMORY SET

WRIST DIRECTION

 INDICATOR

*Note!

 The 3D positioning function is optional. If it is switched off, the device will not show the symbols to 

    indicate the optimal position of the wrist while measuring.

 User can press “START/STOP” button to exit setting mode and go back to standby mode at any step.

*Note!

 When taking repeated measurements, make sure to select the same memory 

    zone number before pressing “START/STOP” button, so the measurement is 

    registered to the appropriate memory.

 Do not move or talk during measurement.

*Note!

 To check the pre-set personalized levels, press and hold "MODE" button.

*Note!

 If there is no data in the selected memory zone, nothing (except month, date, time 

    and memory zone) will appear on the display.

 If there is one data in the selected memory zone, the measurement data will 

    appear on the display.

 If there are 2 data in the selected memory zone, the average of the 2 data will 

    appear on the display.

D. 3D positioning function during measurement

Ǻ

     a. While measuring, slowly move your wrist 

         according to the Arrow symbols on screen 

         to adjust the wrist angle.

     b. When the wrist is in the right position, all Arrow symbols will disappear and PULSE 

         symbol will flash twice on screen. Then the measurement will automatically begin.

A. Press "

" or "

" button to select a particular memory zone (1, 2 or 3).

B. Press "START/STOP" button to start measurement. The monitor will automatically 

     inflate to the level that is right for you.

C. After inflation of the cuff, the pressure will slowly decrease and when a pulse is 

    detected, the SYMBOL will start flashing.

D. When the measurement is finished, the systolic pressure, diastolic pressure, heart rate 

     will be displayed for 1 minute.

E. If the Systolic or Diastolic pressure of the result is greater than your pre-set 

     personalized level, the digit of Systolic or Diastolic pressure will flash. If the Systolic or 

     Diastolic pressure of the result is equal to or less than your pre-set personalized level, 

     the SMILE FACE symbol will flash.

F. The device will switch to standby mode without any operation after 1 minute.

A. Press "MODE" button, the screen displays “3d on” (the device is default with the 

     position sensor function on).

B. Press "

" or "

" button to switch on/off.

C. Press "MODE" button to go back to standby mode.

8-6. Storing, recalling, and erasing measurement data:

8-5. Steps to take blood pressure measurement:

A. Storing data:

After each measurement, the systolic and diastolic pressure, heart rate with the time and 

date will be automatically stored. The monitor features a 3 zone memory capability. Each 

memory zone holds the last 40 measurements, replacing the oldest data with new one.

B. Recalling data:

C. Erasing data:

a.Press "

" and "

" button to select the memory zone.

b. Press "MEMORY" button to enter MEMORY MODE. The average of last 3 measurement

    readings will first appear on the display.

c. Press "MEMORY" button. The latest measurement data will appear with the number as well 

    as the recording date and time. Data are recalled from the latest measurement to the oldest.

d. Press "MEMORY" button to read the next measurement data.

e. To stop reading the memories, press the "START/STOP" button.

      a. Press "

" or "

" button to select the memory zone (1, 2 or 3), then press "MEMORY" button.

      b. Press and hold "MODE" and "

" buttons simultaneously, all stored data, 

          including the average reading, in the selected memory zone will be erased.

      c. To confirm the data has been erased, press "MEMORY" button, no data

          (except month, date, time and memory zone number) should appear.

9. Maintenance

9-1. Clean the device body, case and the cuff carefully with a clean, soft and slightly dampened cloth and 

        then wipe them dry with another piece of dry cloth.

9-2. Do not use thinner, alcohol, petrol (gasoline) or any strong cleansers to clean.

9-3. Do not wash the cuff or make it wet, and never use chemical cleaner on it.

9-4. Do not in any way twist the cuff.

9-5. Make sure no water will enter the main unit.

9-6. When the unit is not to be used for a long time, remove the batteries. (Leaking of battery liquid can 

        cause damage.)

9-7. Do not attempt to disassemble or change any parts of the monitor, including cuff.

9-8. If any suggestion or service is requested please consult your service station.

9-9. Do not place the item directly under the sunlight, in high temperature, or humid and dusty places.

9-10. Do not store in extremely low (less than - 20 °C/ -4 °F) or high (more than 70 °C/ 158 °F) temperature.

Display mark

Corrective action

Battery symbol

Measurement in progress.

Do not move or talk.
Replace all two batteries with new ones. 

Insert the batteries in correct polarities.

Pulse symbol Flashes on the LCD when pulse is

detected during measurement.

Condition/Cause

Appears when the battery voltage is 

excessively low.

Measuring 

error symbol

Press "START/STOP" button again and re-measure. 

Or check cuff if wrapped properly around the wrist 

according to instructions. Check palm if exerting effort. 

Check if there is moving during measurement. Check if 

posture is correct.

Appears when the accurate blood pressure 

could not be obtained.

IHB symbol

Repeat the measurement.

Relax for at least 5 minutes and sit comfortably 

and quietly before you restart a measurement.

Appears for 1 minute when the user was 

talking, moving or shaking or an irregular 

heart beat was detected during measurement.

10. Display Messages/Troubleshooting

Model Number

Method of Measurement

Range of Measurement

Accuracy 

Inflation

Exhaust

Display

Dimension

Unit Weight

Sets of Memory

Cuff Size

Operating Conditions

Storage Conditions

Power Supply

Battery Life

Accessories

BPM280XL

Oscillometric

Pressure 0 ~ 300 mmHg,

Pulse 40 ~ 199 beats/minute

Pr/- 3 mmHg, Pulse +/- 5%

Automatic inflation (Air Pump)

Automatic exhaust valve

Liquid Crystal Display

6.7 x 3.4 x 8.5cm (L x W x H)

Approx. 148 g (including battery)

3 databases of 40 memory each = total 120 sets

Wrist circumference approx. 135 ~ 195 mm

+10 °C to + 40 °C, less than 85% R.H.

- 20 °C to + 70 °C, less than 85% R.H.

AAA (1.5 V) Alkaline battery x 2

Approx. 250 times measurement

Carrying case, instruction manual, 2pcs”AAA” (1.5 V) battery

These specications are subject to change without notice for purpose of improvement.

11. Specifications

0197

This blood pressure monitor complies with the EC Directive and bears the CE 

mark “CE0197”. This blood pressure monitor also complies with mainly 

following standards,(included but not limited) 

Safety standard: 

EN 60601-1 Medical electrical equipment part 1: General requirements for safety

EMC standard: 

EN 60601-1-2 Medical electrical equipment part 1-2: General requirements for safety- Collateral 

                          standard: Electromagnetic compatibility- Requirements and tests

Performance standards: 

EN 1060-1  Non-invasive sphygmomanometers - General requirements

EN 1060-3  Non-invasive sphygmomanometers - Supplementary requirements for 

                     electromechanical blood pressure measuring systems. 

EN 1060-4  Non-invasive sphygmomanometers - Test procedures to determine the overall 

                     system accuracy of automated non-invasive sphygmomanometers.

Note

 To ensure precise measurement reading, 

recalibration of the device is recommended 
after 2 years from the date of purchase. 
Shipping plus handling cost and recalibration 
service fee shall be charged accordingly.

Important /Note! Read the operating 

instructions.

Consult instructions thoroughly before 

use.

Classification:

- Internally powered equipment

- BF type applied part

- IPX0

- Not suitable for use in presence of 

  flammable anaesthetic mixture with 

  air or with Oxygen or nitrous oxide

- Continuous operation with short-time 

  loading

To avoid inaccurate results caused by 

electromagnetic interference between 

electrical and electronic equipments, do 

not use the device near a cell phone or 

microwave oven.

Discard the used product to the recycling 

collection point according to local 

regulations.

Authorized Representative in the 
European Community

EMERGO EUROPE 

Molenstraat 15, 2513 BH, 

The Hague, The Netherlands

TEL : +31-70-3458570     

FAX : +31-70-3467299

Manufacturer:

HEALTH & LIFE Co., Ltd.

9F, No. 186, Jian Yi Road, 

Chung Ho City 235, Taipei, Taiwan

Distributor:

Sabre Nederland b.v.

Ellermanstraat 23 

1099 BX Amsterdam the Netherlands

MODE

START

STOP

MEMORY

* Figure No.6

6. Attaching Pressure Cuff

6-1. Do not place the pressure cuff over a jacket or sweater 
        sleeve. Wrap the pressure cuff around the bare wrist with 
        the monitor facing you.
6-2. Wrap the cuff snugly. Do not make it too tight. 
       Fold the remaining part of the cuff back out of the way.
6-3. Leave approximately 10 mm (Approx. 0.4 inch) between 
        the cuff and the bottom of your hand palm.

*Note!

 Do not use this device if your wrist has any wound or injury.

 Do not wrap the cuff around any body part other than your wrist.

0.4” (10 mm)

135 ~ 195 mm

8-3. Activating Personalized Target Limit Function and Setting Personalized Levels:

8-4. Activating 3D Positioning Technology Function:

*Note!

 Systolic pressure range for setting personalized levels is between 70 and 250 mmHg.

 Diastolic pressure range for setting personalized levels is between 40 and 115 mmHg.

 The default value for Systolic pressure is 135 mmHg and 85 mmHg for Diastolic pressure.

A. Press "MODE" button (" - - " will flash for 1st systolic pre-set value).

     Press "

" or "

" button to set/adjust the personalized systolic pressure.

B. Press "MODE" button (" - - " will flash for 1st diastolic pre-set value).

     Press "

" or "

" button to set/adjust the personalized diastolic pressure.

C. Press "MODE" button (" - - " will flash for 2nd systolic pre-set value).

     Press "

" or "

" button to set/adjust the personalized systolic pressure.

D. Press "MODE" button (" - - " will flash for 2nd diastolic pre-set value).

     Press "

" or "

" button to set/adjust the personalized diastolic pressure.

E. Press "MODE" button (" - - " will flash for 3rd systolic pre-set value).

     Press "

" or "

" button to set/adjust the personalized systolic pressure.

F. Press "MODE" button (" - - " will flash for 3rd diastolic pre-set value).

     Press "

" or "

" button to set/adjust the personalized diastolic pressure.

G. After setting the personalized levels, the device will return to Standby Mode and is 

     ready for a measurement by pressing START/STOP button.

* Figure No.7

7-1. To get accurate measurements, make sure the cuff is at the same height as the heart. 

        Having the cuff higher or lower may cause inaccurate results.

7-2. Measurement while sitting down:

A. Place your elbow on a table or other object (such as the carrying case).

B. Use the armrest to position the arm so that the pressure cuff is at the same 

     height as the heart.

C. Relax your hand with the palm facing up.

7. Correct Measurement Position

8-1. Inserting batteries in position:

8. How to Measure

A. Open battery cover.

B. Refer to the picture in positioning batteries. (Figure No.5)

C. All segments will appear on the display in 3 seconds.

8-2. Setting Date and Time:

A. Press "MODE" button ("month" starts flashing). Press "

" or "

" button to set the correct month. 

B. Press "MODE" button again ("date" starts flashing). Press "

" or "

" button to set the correct date.

C. Press "MODE" button again ("hour" starts flashing). Press "

" or "

" button to set the correct hour 

     in 12-hour format.

D. Press "MODE" button again ("minute" starts flashing). Press "

" or "

" button to set the correct 

     minute.

E. Press "MODE" button again to switch to normal time.

When the low battery symbol               appears on the display, please change the batteries.

Replace all the batteries with new ones and do not mix new and old batteries. Do not mix alkaline, 

standard (carbon-zinc) or rechargeable (cadmium) 

batteries either. It may shorten the battery life or cause 

the device to malfunction. Remove the battery cover at 

the bottom of the unit and insert 2 pieces AAA alkaline 

batteries into the battery compartment. Make sure the 

polarities "+" and "–" ends are properly positioned.

5. Inserting/ Changing Batteries

* Figure No.5

Move wrist upward

Move wrist right

Move wrist left 

Move wrist downward 

P/N: 323101293    VER:001